Critical warnings from b-files' comments for days between 2019299 and 2019299 (0 sec, 3 rows)

TypeCommentcountpriorityfnamelocationserialtimevalueminmax
rl:2019 298 05 15 25 Motor #2: appears stuck with a reference sensor activated.21B29919.224TORON#22405:09:121-1
rl:2019 298 05 41 14 New value set for watermark LOOP.RUN.TIME: 653.999996 millisec.63B29919.224TORON#22405:09:261-1
rl:2019 298 05 41 14 New value set for watermark LOOP.RUN.TIME: 653.999996 millisec.63B29919.224TORON#22405:09:261-1

Instrument constant changes for the week before 2019299 (3 sec, 0 rows)

SL burned out? (AP results when SL is ON) (0 sec, 24 rows)

SerialYYYYJJJSL CurrentSL Voltage
2242019294-12.1123.84
2242019295-12.1123.84
2242019296-12.1123.81
2242019297-12.123.84
2242019298-12.123.84
2242019299-12.123.84

Non-critical warnings from b-files' comments for days between 2019299 and 2019299 (0 sec, 5 rows)

TypeCommentcountpriorityfnamelocationserialtimevalueminmax
mf:Illegal command (mf) in schedule ecsc, ignoring199B29919.224TORON#22402:11:591-1
rl:2019 298 05 14 15 Reset Warm -> Tepid from 2019 298 05 14 09 until 2019 298 05 14 09 . Counts: 1,013B29919.224TORON#22405:09:091-1
rl:2019 298 05 15 43 <M,6,242>23B29919.224TORON#22405:09:151-1
rl:2019 298 05 15 43 <M,6,242>25B29919.224TORON#22405:09:151-1
ds:DS intensity too high ( 212334/cy) quitting.399B29919.224TORON#22416:18:491-1

Measurements and tests summary.


_jday_loc_serial_skcdsdtedfzhghprsslzs
299TORONTO#224ecsc,Next schedule will be ecsc (now ecsc ZS= 93 ZS= 93.35)7521713122646
(0 sec)

Number of days since last dome/window cleaning (0 sec, 2 rows)

serialdays
224318

Missing files (0 sec, 1 rows)

Seriallast datadays ago
22420193000-1

Run/Stop results for last week. (0 sec, 1 rows)

SerialTotalFailed 0Failed darkFailed 1Failed 2Failed 3Failed 4Failed 5Failed double
2241503000000

AP results for MB Brewer. (0 sec, 1 rows)

SerialJdayt1 offt1 ont2 offt2 ont3 offt3 onHV offHV on+15V off+15V on+5V off+5V on-15V off-15V on+24V off+24V on+5V sps off+5V sps on-8V sps off-8V sps onSL A offSL A onSL V offSL V onW
22420192999.80.010.10.09.11.31349.32.212.05.95.10.6-12.12.024.62.35.012.0-5.05.10.0-12.10.023.83-288.5

Standard lamp intensity analysis (0 sec, 1 rows)

Serialminmaxfilterfixable30-day history
#224313214031336300yes