Critical warnings from b-files' comments for days between 2019190 and 2019190 (0 sec, 3 rows)

TypeCommentcountpriorityfnamelocationserialtimevalueminmax
rl:2019 189 05 38 51 Motor #2: appears stuck with a reference sensor activated.21B19019.224TORON#22405:29:551-1
rl:2019 189 06 00 05 New value set for watermark LOOP.RUN.TIME: 653.999996 millisec.93B19019.224TORON#22405:29:591-1
rl:2019 189 06 00 05 New value set for watermark LOOP.RUN.TIME: 653.999996 millisec.93B19019.224TORON#22405:29:591-1

Instrument constant changes for the week before 2019190 (3 sec, 0 rows)

SL burned out? (AP results when SL is ON) (0 sec, 21 rows)

SerialYYYYJJJSL CurrentSL Voltage
2242019183-12.1423.9
2242019184-12.1423.9
2242019185-12.1423.93
2242019186-12.1423.87
2242019187-12.1523.87
2242019188-12.1423.87
2242019189-12.1323.87
2242019190-12.1323.87

Hg with step corrections larger than 10 steps or multiple corrections larger than 2.5 steps (0 sec, 1 rows)

SerialYYYYJJJHG largemax correction
2242019190215.6

Non-critical warnings from b-files' comments for days between 2019190 and 2019190 (0 sec, 3 rows)

TypeCommentcountpriorityfnamelocationserialtimevalueminmax
rl:2019 189 05 37 41 Reset Warm -> Tepid from 2019 189 05 37 35 until 2019 189 05 37 35 . Counts: 1,013B19019.224TORON#22405:29:521-1
rl:2019 189 06 00 05 New value set for watermark LOOP.RUN.TIME: 653.999996 millisec.43B19019.224TORON#22405:29:591-1
rl:2019 189 06 00 05 New value set for watermark LOOP.RUN.TIME: 653.999996 millisec.43B19019.224TORON#22405:29:591-1

Measurements and tests summary.


_jday_loc_serial_skcdsdtedfmfzhghprssluvzs
190TORONTO#224calsc9021137353127287
(0 sec)

New operators? (0 sec, 85 rows)

Number of days since last dome/window cleaning (0 sec, 2 rows)

serialdays
224209

Missing files (0 sec, 1 rows)

Seriallast datadays ago
22420191910-1

Run/Stop results for last week. (0 sec, 1 rows)

SerialTotalFailed 0Failed darkFailed 1Failed 2Failed 3Failed 4Failed 5Failed double
2241603000000

AP results for MB Brewer. (0 sec, 1 rows)

SerialJdayt1 offt1 ont2 offt2 ont3 offt3 onHV offHV on+15V off+15V on+5V off+5V on-15V off-15V on+24V off+24V on+5V sps off+5V sps on-8V sps off-8V sps onSL A offSL A onSL V offSL V onW
224201919027.30.027.40.027.21.41346.42.412.03.35.10.9-12.12.024.62.35.012.0-5.05.10.0-12.10.023.91-289.5

Standard lamp intensity analysis (0 sec, 1 rows)

Serialminmaxfilterfixable30-day history
#224299551029969500yes