Database last updated on Sat, 12 Sep 2026 14:40:53 +0000 by localhost

Critical warnings from b-files' comments for days between 2019144 and 2019144 (0 sec, 3 rows)

TypeCommentcountpriorityfnamelocationserialtimevalueminmax
rl:2019 143 05 26 07 Motor #2: appears stuck with a reference sensor activated.21B14419.224TORON#22405:21:491-1
rl:2019 143 05 47 42 New value set for watermark LOOP.RUN.TIME: 653.999996 millisec.73B14419.224TORON#22405:21:541-1
rl:2019 143 05 47 42 New value set for watermark LOOP.RUN.TIME: 653.999996 millisec.73B14419.224TORON#22405:21:541-1

Instrument constant changes for the week before 2019144 (3 sec, 0 rows)

SL burned out? (AP results when SL is ON) (0 sec, 28 rows)

SerialYYYYJJJSL CurrentSL Voltage
2242019137-12.1124.08
2242019138-12.1123.93
2242019139-12.1123.9
2242019140-12.1323.9
2242019141-12.123.9
2242019142-12.1123.9
2242019143-12.1123.9
2242019144-12.1124.05

Hg with step corrections larger than 10 steps or multiple corrections larger than 2.5 steps (0 sec, 1 rows)

SerialYYYYJJJHG largemax correction
2242019144111.6

Non-critical warnings from b-files' comments for days between 2019144 and 2019144 (0 sec, 3 rows)

TypeCommentcountpriorityfnamelocationserialtimevalueminmax
rl:2019 143 05 26 03 Reset Warm -> Tepid from 2019 143 05 25 58 until 2019 143 05 25 58 . Counts: 1,013B14419.224TORON#22405:21:471-1
ds:Suppressed oscillation of filter wheel.45B14419.224TORON#22414:29:371-1
sc:SC intensity too low ( 408/cy), skipping.23B14419.224TORON#22422:32:051-1

Measurements and tests summary.


_jday_loc_serial_skcdsdtedfmfzhghprssluvzs
144TORONTO#224calsc8121853534252913
(0 sec)

Number of days since last dome/window cleaning (0 sec, 2 rows)

serialdays
224163

Missing files (0 sec, 1 rows)

Seriallast datadays ago
22420191450-1

Run/Stop results for last week. (0 sec, 1 rows)

SerialTotalFailed 0Failed darkFailed 1Failed 2Failed 3Failed 4Failed 5Failed double
2241605000000

AP results for MB Brewer. (0 sec, 1 rows)

SerialJdayt1 offt1 ont2 offt2 ont3 offt3 onHV offHV on+15V off+15V on+5V off+5V on-15V off-15V on+24V off+24V on+5V sps off+5V sps on-8V sps off-8V sps onSL A offSL A onSL V offSL V onW
224201914422.40.022.50.021.21.41347.12.412.07.75.11.6-12.12.024.62.35.012.0-5.05.10.0-12.10.023.95-291.2

Standard lamp intensity analysis (0 sec, 1 rows)

Serialminmaxfilterfixable30-day history
#224296123029668500yes