Database last updated on Sun, 13 Sep 2026 12:41:32 +0000 by localhost
Critical warnings from b-files' comments for days between 2019128 and 2019128 (0 sec, 3 rows)
| Type | Comment | count | priority | fname | location | serial | time | value | min | max |
| rl: | 2019 127 05 26 29 Motor #2: appears stuck with a reference sensor activated. | 2 | 1 | B12819.224 | TORON | #224 | 05:18:30 | | 1 | -1 |
| rl: | 2019 127 05 52 33 New value set for watermark LOOP.RUN.TIME: 699.000025 millisec. | 5 | 3 | B12819.224 | TORON | #224 | 05:18:34 | | 1 | -1 |
| rl: | 2019 127 05 52 33 New value set for watermark LOOP.RUN.TIME: 699.000025 millisec. | 5 | 3 | B12819.224 | TORON | #224 | 05:18:34 | | 1 | -1 |
Instrument constant changes for the week before 2019128 (1 sec, 0 rows)
SL burned out? (AP results when SL is ON) (0 sec, 32 rows)
| Serial | YYYYJJJ | SL Current | SL Voltage |
| 224 | 2019121 | -12.1 | 23.87 |
| 224 | 2019122 | -12.11 | 24.02 |
| 224 | 2019123 | -12.1 | 23.9 |
| 224 | 2019124 | -12.11 | 23.9 |
| 224 | 2019125 | -12.11 | 23.93 |
| 224 | 2019126 | -12.11 | 23.9 |
| 224 | 2019127 | -12.11 | 23.9 |
| 224 | 2019128 | -12.1 | 23.9 |
HG burned out? (AP results when HG is ON) (0 sec, 81 rows)
Non-critical warnings from b-files' comments for days between 2019128 and 2019128 (1 sec, 1 rows)
| Type | Comment | count | priority | fname | location | serial | time | value | min | max |
| rl: | 2019 127 05 25 19 Reset Warm -> Tepid from 2019 127 05 25 13 until 2019 127 05 25 14 . Counts: 1,0 | 1 | 3 | B12819.224 | TORON | #224 | 05:18:27 | | 1 | -1 |
Measurements and tests summary.
| _jday | _loc | _serial | _skc | ds | dt | ed | fz | hg | hp | rs | sl | uv | zs |
| 128 | TORONTO | #224 | calsc | 93 | 2 | 1 | 2 | 30 | 30 | 2 | 7 | 26 | 7 |
(1 sec)
New operators? (0 sec, 62 rows)
Number of days since last dome/window cleaning (0 sec, 2 rows)
Missing files (0 sec, 1 rows)
| Serial | last data | days ago |
| 224 | 2019129 | 0-1 |
Run/Stop results for last week. (0 sec, 1 rows)
| Serial | Total | Failed 0 | Failed dark | Failed 1 | Failed 2 | Failed 3 | Failed 4 | Failed 5 | Failed double |
| 224 | 16 | 0 | 8 | 0 | 0 | 0 | 0 | 0 | 0 |
AP results for MB Brewer. (0 sec, 1 rows)
| Serial | Jday | t1 off | t1 on | t2 off | t2 on | t3 off | t3 on | HV off | HV on | +15V off | +15V on | +5V off | +5V on | -15V off | -15V on | +24V off | +24V on | +5V sps off | +5V sps on | -8V sps off | -8V sps on | SL A off | SL A on | SL V off | SL V on | W |
| 224 | 2019128 | 13.0 | 0.0 | 12.9 | 0.0 | 11.8 | 1.3 | 1349.1 | 2.2 | 12.0 | 6.9 | 5.1 | 0.8 | -12.1 | 2.0 | 24.7 | 2.3 | 5.0 | 12.0 | -5.0 | 5.1 | 0.0 | -12.1 | 0.0 | 23.97 | -289.2 |
Standard lamp intensity analysis (0 sec, 1 rows)
| Serial | min | max | filter | fixable | 30-day history |
| #224 | 2971180 | 2973370 | 0 | yes |