Critical warnings from b-files' comments for days between 2019124 and 2019124 (0 sec, 1 rows)

TypeCommentcountpriorityfnamelocationserialtimevalueminmax
rl:2019 123 05 23 03 Motor #2: appears stuck with a reference sensor activated.21B12419.224TORON#22405:16:571-1

Instrument constant changes for the week before 2019124 (1 sec, 0 rows)

SL burned out? (AP results when SL is ON) (0 sec, 32 rows)

SerialYYYYJJJSL CurrentSL Voltage
2242019117-12.0823.9
2242019118-12.123.87
2242019119-12.0823.87
2242019120-12.123.87
2242019121-12.123.87
2242019122-12.1124.02
2242019123-12.123.9
2242019124-12.1123.9

Non-critical warnings from b-files' comments for days between 2019124 and 2019124 (1 sec, 6 rows)

TypeCommentcountpriorityfnamelocationserialtimevalueminmax
rl:2019 123 05 21 55 Reset Warm -> Tepid from 2019 123 05 21 50 until 2019 123 05 21 50 . Counts: 1,013B12419.224TORON#22405:16:541-1
rl:2019 123 05 44 00 New value set for watermark LOOP.RUN.TIME: 711.999989 millisec.13B12419.224TORON#22405:17:011-1
rl:2019 123 05 44 00 New value set for watermark LOOP.RUN.TIME: 711.999989 millisec.13B12419.224TORON#22405:17:011-1
rl:2019 124 11 53 06 New value set for watermark LOOP.RUN.TIME: 714.000034 millisec.23B12419.224TORON#22405:17:041-1
rl:2019 124 11 53 06 New value set for watermark LOOP.RUN.TIME: 714.000034 millisec.23B12419.224TORON#22405:17:041-1
zs:ZS intensity too low ( 141/cy), skipping.299B12419.224TORON#22410:12:141-1

Measurements and tests summary.


_jday_loc_serial_skcdsdtedfzhghprssluvzs
124TORONTO#224ecsc7121398245522
(1 sec)

Number of days since last dome/window cleaning (0 sec, 2 rows)

serialdays
224143

Missing files (0 sec, 1 rows)

Seriallast datadays ago
22420191250-1

Run/Stop results for last week. (0 sec, 1 rows)

SerialTotalFailed 0Failed darkFailed 1Failed 2Failed 3Failed 4Failed 5Failed double
2241608000000

AP results for MB Brewer. (0 sec, 1 rows)

SerialJdayt1 offt1 ont2 offt2 ont3 offt3 onHV offHV on+15V off+15V on+5V off+5V on-15V off-15V on+24V off+24V on+5V sps off+5V sps on-8V sps off-8V sps onSL A offSL A onSL V offSL V onW
224201912412.90.012.90.012.41.31349.02.312.07.45.10.8-12.12.024.62.35.012.0-5.05.10.0-12.10.023.89-289.4

Standard lamp intensity analysis (0 sec, 1 rows)

Serialminmaxfilterfixable30-day history
#224296246029659400yes