Critical warnings from b-files' comments for days between 2019114 and 2019114 (0 sec, 1 rows)

TypeCommentcountpriorityfnamelocationserialtimevalueminmax
rl:2019 113 05 25 37 Motor #2: appears stuck with a reference sensor activated.21B11419.224TORON#22405:19:441-1

Instrument constant changes for the week before 2019114 (2 sec, 0 rows)

SL burned out? (AP results when SL is ON) (0 sec, 27 rows)

SerialYYYYJJJSL CurrentSL Voltage
2242019107-12.123.87
2242019108-12.123.87
2242019109-12.1323.87
2242019110-12.123.87
2242019111-12.1123.87
2242019112-12.1123.87
2242019113-12.1123.87
2242019114-12.123.87

Non-critical warnings from b-files' comments for days between 2019114 and 2019114 (1 sec, 4 rows)

TypeCommentcountpriorityfnamelocationserialtimevalueminmax
rl:2019 113 05 24 27 Reset Warm -> Tepid from 2019 113 05 24 21 until 2019 113 05 24 21 . Counts: 1,013B11419.224TORON#22405:19:421-1
rl:2019 113 05 42 27 New value set for watermark LOOP.RUN.TIME: 673.000050 millisec.43B11419.224TORON#22405:19:491-1
rl:2019 113 05 42 27 New value set for watermark LOOP.RUN.TIME: 673.000050 millisec.43B11419.224TORON#22405:19:491-1
zs:ZS intensity too low ( 181/cy), skipping.499B11419.224TORON#22410:23:061-1

Measurements and tests summary.


_jday_loc_serial_skcdsdtedfmfzhghprssluvzs
114TORONTO#224ecsc60211111312245623
(1 sec)

Number of days since last dome/window cleaning (0 sec, 2 rows)

serialdays
224133

Missing files (0 sec, 1 rows)

Seriallast datadays ago
22420191150-1

Run/Stop results for last week. (0 sec, 1 rows)

SerialTotalFailed 0Failed darkFailed 1Failed 2Failed 3Failed 4Failed 5Failed double
2241604000000

AP results for MB Brewer. (0 sec, 1 rows)

SerialJdayt1 offt1 ont2 offt2 ont3 offt3 onHV offHV on+15V off+15V on+5V off+5V on-15V off-15V on+24V off+24V on+5V sps off+5V sps on-8V sps off-8V sps onSL A offSL A onSL V offSL V onW
224201911410.50.010.60.09.51.31349.42.212.06.95.10.7-12.12.024.62.25.012.0-5.05.10.0-12.10.023.87-288.8

Standard lamp intensity analysis (0 sec, 1 rows)

Serialminmaxfilterfixable30-day history
#224296293029684100yes