Database last updated on Sat, 12 Sep 2026 22:11:02 +0000 by localhost

Critical warnings from b-files' comments for days between 2019094 and 2019094 (0 sec, 3 rows)

TypeCommentcountpriorityfnamelocationserialtimevalueminmax
rl:2019 093 05 31 41 Motor #2: appears stuck with a reference sensor activated.21B09419.224TORON#22405:25:531-1
rl:2019 093 05 57 13 New value set for watermark LOOP.RUN.TIME: 685.999966 millisec.63B09419.224TORON#22405:25:571-1
rl:2019 093 05 57 13 New value set for watermark LOOP.RUN.TIME: 685.999966 millisec.63B09419.224TORON#22405:25:571-1

Instrument constant changes for the week before 2019094 (3 sec, 0 rows)

SL burned out? (AP results when SL is ON) (0 sec, 24 rows)

SerialYYYYJJJSL CurrentSL Voltage
2242019087-12.0823.93
2242019088-12.123.93
2242019089-12.123.93
2242019090-12.0823.93
2242019091-12.0823.9
2242019092-12.0823.93
2242019093-12.123.9
2242019094-12.0823.9

Non-critical warnings from b-files' comments for days between 2019094 and 2019094 (0 sec, 4 rows)

TypeCommentcountpriorityfnamelocationserialtimevalueminmax
sc:SC intensity too low ( 2/cy), skipping.13B09419.224TORON#22403:31:471-1
rl:2019 093 05 30 30 Reset Warm -> Tepid from 2019 093 05 30 24 until 2019 093 05 30 25 . Counts: 1,013B09419.224TORON#22405:25:501-1
ds:DS intensity too low ( 7/cy), skipping.399B09419.224TORON#22411:09:461-1
zs:ZS intensity too low ( 311/cy), skipping.199B09419.224TORON#22411:09:591-1

Measurements and tests summary.


_jday_loc_serial_skcdsdtedfzhghprssluvzs
094TORONTO#224ecsc642111110255216
(0 sec)

Number of days since last dome/window cleaning (0 sec, 2 rows)

serialdays
224113

Missing files (0 sec, 1 rows)

Seriallast datadays ago
22420190950-1

Run/Stop results for last week. (0 sec, 1 rows)

SerialTotalFailed 0Failed darkFailed 1Failed 2Failed 3Failed 4Failed 5Failed double
2241608000000

AP results for MB Brewer. (0 sec, 1 rows)

SerialJdayt1 offt1 ont2 offt2 ont3 offt3 onHV offHV on+15V off+15V on+5V off+5V on-15V off-15V on+24V off+24V on+5V sps off+5V sps on-8V sps off-8V sps onSL A offSL A onSL V offSL V onW
22420190947.20.07.20.05.61.31350.02.112.05.25.10.4-12.12.024.62.25.012.0-5.05.10.0-12.10.023.89-288.7

Standard lamp intensity analysis (0 sec, 1 rows)

Serialminmaxfilterfixable30-day history
#224296370029640600yes