Database last updated on Sun, 13 Sep 2026 14:11:09 +0000 by localhost

Critical warnings from b-files' comments for days between 2019070 and 2019070 (1 sec, 3 rows)

TypeCommentcountpriorityfnamelocationserialtimevalueminmax
rl:2019 069 05 38 10 Motor #2: appears stuck with a reference sensor activated.31B07019.224TORON#22405:35:081-1
rl:2019 069 06 04 17 New value set for watermark LOOP.RUN.TIME: 683.000040 millisec.63B07019.224TORON#22405:35:131-1
rl:2019 069 06 04 17 New value set for watermark LOOP.RUN.TIME: 683.000040 millisec.63B07019.224TORON#22405:35:131-1

Instrument constant changes for the week before 2019070 (1 sec, 0 rows)

SL burned out? (AP results when SL is ON) (0 sec, 24 rows)

SerialYYYYJJJSL CurrentSL Voltage
2242019063-12.0823.9
2242019064-12.0823.76
2242019065-12.0823.9
2242019066-12.0823.96
2242019067-12.0823.93
2242019068-12.123.96
2242019069-12.123.96
2242019070-12.123.9

Non-critical warnings from b-files' comments for days between 2019070 and 2019070 (0 sec, 5 rows)

TypeCommentcountpriorityfnamelocationserialtimevalueminmax
rl:2019 069 05 37 04 Reset Warm -> Tepid from 2019 069 05 36 58 until 2019 069 05 36 58 . Counts: 1,013B07019.224TORON#22405:35:061-1
rl:2019 069 06 04 17 New value set for watermark LOOP.RUN.TIME: 683.000040 millisec.43B07019.224TORON#22405:35:131-1
rl:2019 069 06 04 17 New value set for watermark LOOP.RUN.TIME: 683.000040 millisec.43B07019.224TORON#22405:35:131-1
rl:2019 070 06 40 06 New value set for watermark LOOP.RUN.TIME: 716.999960 millisec.23B07019.224TORON#22405:35:251-1
rl:2019 070 06 40 06 New value set for watermark LOOP.RUN.TIME: 716.999960 millisec.23B07019.224TORON#22405:35:251-1

Measurements and tests summary.


_jday_loc_serial_skcdsdtedhghprssluvzs
070TORONTO#224ecsc42211211264431
(1 sec)

New operators? (0 sec, 91 rows)

Number of days since last dome/window cleaning (0 sec, 2 rows)

serialdays
224089

Missing files (0 sec, 1 rows)

Seriallast datadays ago
22420190710-1

Run/Stop results for last week. (0 sec, 1 rows)

SerialTotalFailed 0Failed darkFailed 1Failed 2Failed 3Failed 4Failed 5Failed double
2241409000000

AP results for MB Brewer. (0 sec, 1 rows)

SerialJdayt1 offt1 ont2 offt2 ont3 offt3 onHV offHV on+15V off+15V on+5V off+5V on-15V off-15V on+24V off+24V on+5V sps off+5V sps on-8V sps off-8V sps onSL A offSL A onSL V offSL V onW
22420190707.50.07.60.06.71.31349.42.112.05.15.10.4-12.12.024.72.25.012.0-5.05.10.0-12.10.023.90-289.2

Standard lamp intensity analysis (0 sec, 1 rows)

Serialminmaxfilterfixable30-day history
#224295096029531600yes