Critical warnings from b-files' comments for days between 2023270 and 2023270 (0 sec, 4 rows)

TypeCommentcountpriorityfnamelocationserialtimevalueminmax
rl:2023 269 05 57 04 Motor #2: appears stuck with a reference sensor activated.11#22305:44:111-1
rl:2023 269 06 13 40 New value set for watermark LOOP.RUN.TIME: 700.000000 millisec.63#22305:44:141-1
rl:2023 269 06 13 40 New value set for watermark LOOP.RUN.TIME: 700.000000 millisec.63#22305:44:141-1
rl:2023 270 18 02 37 PMT counter failure11#22305:44:301-1

Instrument constant changes for the week before 2023270 (2 sec, 0 rows)

HG burned out? (AP results when HG is ON) (0 sec, 15 rows)

Non-critical warnings from b-files' comments for days between 2023270 and 2023270 (0 sec, 7 rows)

TypeCommentcountpriorityfnamelocationserialtimevalueminmax
rl:2023 269 05 55 47 Bad command string. Command string buffer cleared.13#22305:44:061-1
rl:2023 269 05 55 47 Bad command string. Command string buffer cleared.13#22305:44:061-1
rl:2023 269 05 55 47 <_13#22305:44:081-1
rl:2023 269 06 13 40 New value set for watermark LOOP.RUN.TIME: 700.000000 millisec.33#22305:44:141-1
rl:2023 269 06 13 40 New value set for watermark LOOP.RUN.TIME: 700.000000 millisec.33#22305:44:141-1
rl:2023 270 08 04 07 New value set for watermark LOOP.RUN.TIME: 715.000010 millisec.33#22305:44:221-1
rl:2023 270 08 04 07 New value set for watermark LOOP.RUN.TIME: 715.000010 millisec.33#22305:44:221-1

Measurements and tests summary.


_jday_loc_serial_skcaodedsdtedfzhghprsslzs
270EUREKA PEARL#223calds,Next schedule will be calds (now calds ZS= 93 ZS= 93.25599),Waiting until ZA =22225819182774
(1 sec)

Number of days since last dome/window cleaning (0 sec, 3 rows)

serialdays
223more than a year
3more than a year

Run/Stop results for last week. (0 sec, 1 rows)

SerialTotalFailed 0Failed darkFailed 1Failed 2Failed 3Failed 4Failed 5Failed double
2231603000000

AP results for SB Brewers. (0 sec, 1 rows)

SerialJdayt1 offt1 ont2 offt2 ont3 offt3 onHV offHV on+12V off+12V on+5V off+5V on-12V off-12V on+24V off+24V on+5V sps off+5V sps onSL A offSL A onSL V offSL V onWatts
223202327017.417.416.616.614.514.51261.21261.412.012.05.05.0-12.0-12.024.024.05.05.00.01.50.010.3115.8

Standard lamp intensity analysis (0 sec, 1 rows)

Serialminmaxfilterfixable30-day history
#223267992026864000yes