Database last updated on Wed, 02 Sep 2026 23:42:08 +0000 by localhost

Critical warnings from b-files' comments for days between 2023243 and 2023243 (1 sec, 4 rows)

TypeCommentcountpriorityfnamelocationserialtimevalueminmax
rl:2023 242 06 04 21 Motor #2: appears stuck with a reference sensor activated.11#22305:53:081-1
rl:2023 242 06 20 46 New value set for watermark LOOP.RUN.TIME: 653.999996 millisec.53#22305:53:111-1
rl:2023 242 06 20 46 New value set for watermark LOOP.RUN.TIME: 653.999996 millisec.53#22305:53:111-1
rl:2023 243 07 05 03 PMT counter failure11#22305:53:231-1

Instrument constant changes for the week before 2023243 (1 sec, 0 rows)

HG burned out? (AP results when HG is ON) (0 sec, 23 rows)

Non-critical warnings from b-files' comments for days between 2023243 and 2023243 (0 sec, 3 rows)

TypeCommentcountpriorityfnamelocationserialtimevalueminmax
rl:2023 242 06 04 08 Bad command string. Command string buffer cleared.23#22305:52:591-1
rl:2023 242 06 04 08 Bad command string. Command string buffer cleared.23#22305:52:591-1
rl:2023 242 06 04 08 <23#22305:53:011-1

Measurements and tests summary.


_jday_loc_serial_skcaodedsdtedfzhghprsslzs
243EUREKA PEARL#223calds18181258151418152
(1 sec)

Number of days since last dome/window cleaning (0 sec, 3 rows)

serialdays
223more than a year
3more than a year

Run/Stop results for last week. (0 sec, 1 rows)

SerialTotalFailed 0Failed darkFailed 1Failed 2Failed 3Failed 4Failed 5Failed double
223700000000

AP results for SB Brewers. (0 sec, 1 rows)

SerialJdayt1 offt1 ont2 offt2 ont3 offt3 onHV offHV on+12V off+12V on+5V off+5V on-12V off-12V on+24V off+24V on+5V sps off+5V sps onSL A offSL A onSL V offSL V onWatts
223202324317.918.117.317.316.316.31261.11261.412.012.05.05.0-12.0-12.024.024.05.05.00.01.50.010.2915.8

Standard lamp intensity analysis (0 sec, 1 rows)

Serialminmaxfilterfixable30-day history
#223272529027252900yes