Database last updated on Tue, 03 Feb 2026 19:10:52 +0000 by localhost

Critical warnings from b-files' comments for days between 2023232 and 2023232 (0 sec, 4 rows)

TypeCommentcountpriorityfnamelocationserialtimevalueminmax
rl:2023 231 06 05 05 Motor #2: appears stuck with a reference sensor activated.11#22305:57:301-1
rl:2023 231 06 30 57 New value set for watermark LOOP.RUN.TIME: 683.000040 millisec.53#22305:57:331-1
rl:2023 231 06 30 57 New value set for watermark LOOP.RUN.TIME: 683.000040 millisec.53#22305:57:331-1
rl:2023 232 16 51 55 PMT counter failure11#22305:57:451-1

Instrument constant changes for the week before 2023232 (2 sec, 0 rows)

SL burned out? (AP results when SL is ON) (0 sec, 1 rows)

HG burned out? (AP results when HG is ON) (0 sec, 22 rows)

Non-critical warnings from b-files' comments for days between 2023232 and 2023232 (0 sec, 3 rows)

TypeCommentcountpriorityfnamelocationserialtimevalueminmax
rl:2023 231 06 04 53 Bad command string. Command string buffer cleared.13#22305:57:251-1
rl:2023 231 06 04 53 Bad command string. Command string buffer cleared.13#22305:57:251-1
rl:2023 231 06 04 53 <13#22305:57:271-1

Measurements and tests summary.


_jday_loc_serial_skcaodedsdtedfzhghprsslzs
232EUREKA PEARL#223calds34341225121118189
(1 sec)

Number of days since last dome/window cleaning (0 sec, 3 rows)

serialdays
223more than a year
3more than a year

Run/Stop results for last week. (0 sec, 1 rows)

SerialTotalFailed 0Failed darkFailed 1Failed 2Failed 3Failed 4Failed 5Failed double
223805000000

AP results for SB Brewers. (0 sec, 1 rows)

SerialJdayt1 offt1 ont2 offt2 ont3 offt3 onHV offHV on+12V off+12V on+5V off+5V on-12V off-12V on+24V off+24V on+5V sps off+5V sps onSL A offSL A onSL V offSL V onWatts
223202323218.218.317.817.816.716.71261.61261.612.012.05.05.0-12.0-12.024.024.05.05.00.01.50.010.3015.8

Standard lamp intensity analysis (0 sec, 1 rows)

Serialminmaxfilterfixable30-day history
#223273555027355500yes