Database last updated on Wed, 09 Sep 2026 01:40:56 +0000 by localhost

Critical warnings from b-files' comments for days between 2020070 and 2020070 (0 sec, 2 rows)

TypeCommentcountpriorityfnamelocationserialtimevalueminmax
rl:2020 069 06 12 45 Motor #2: appears stuck with a reference sensor activated.41B07020.223EUREK#22306:02:511-1
rl:2020 070 17 33 15 PMT counter failure11B07020.223EUREK#22306:03:071-1

Instrument constant changes for the week before 2020070 (3 sec, 0 rows)

SL burned out? (AP results when SL is ON) (0 sec, 25 rows)

SerialYYYYJJJSL CurrentSL Voltage
2232020063-12.1424.43
2232020064-12.1524.4
2232020065-12.1524.4
2232020066-12.1424.43
2232020067-12.1424.43
2232020068-12.1524.4
2232020069-12.1424.43
2232020070-12.1524.43

Non-critical warnings from b-files' comments for days between 2020070 and 2020070 (0 sec, 8 rows)

TypeCommentcountpriorityfnamelocationserialtimevalueminmax
rl:2020 069 06 10 31 Reset Warm -> Tepid from 2020 069 06 10 25 until 2020 069 06 10 26 . Counts: 1,013B07020.223EUREK#22306:02:491-1
rl:2020 069 06 34 56 New value set for watermark LOOP.RUN.TIME: 700.000000 millisec.33B07020.223EUREK#22306:02:561-1
rl:2020 069 06 34 56 New value set for watermark LOOP.RUN.TIME: 700.000000 millisec.33B07020.223EUREK#22306:02:561-1
cfg_out:Illegal command (cfg199B07020.223EUREK#22306:03:141-1
ux:HOME key pressed199B07020.223EUREK#22316:26:311-1
ux:UV extended scan aborted13B07020.223EUREK#22316:26:331-1
ux:Going to menu199B07020.223EUREK#22316:28:111-1
si:New SI, HC = 4, NC = 475 Old = 4 , 475199B07020.223EUREK#22316:41:571-1

Measurements and tests summary.


_jday_loc_serial_skcaodedsdtedfzhghprssluvzs
070EUREKA PEARL#223ecsc191921181211254526
(0 sec)

Relative and absolute humidity for day 2020070 (2 sec, 1 rows)

SerialRH, %AH, g/m37-day RH historyDes-nt replcd days ago
2235.60.8