Database last updated on Thu, 10 Sep 2026 21:40:59 +0000 by localhost

Critical warnings from b-files' comments for days between 2020059 and 2020059 (0 sec, 4 rows)

TypeCommentcountpriorityfnamelocationserialtimevalueminmax
rl:2020 058 06 13 43 Motor #2: appears stuck with a reference sensor activated.21B05920.223EUREK#22306:06:171-1
rl:2020 058 06 35 13 New value set for watermark LOOP.RUN.TIME: 700.000000 millisec.53B05920.223EUREK#22306:06:221-1
rl:2020 058 06 35 13 New value set for watermark LOOP.RUN.TIME: 700.000000 millisec.53B05920.223EUREK#22306:06:221-1
rl:2020 058 07 20 41 PMT counter failure11B05920.223EUREK#22306:06:311-1

Instrument constant changes for the week before 2020059 (5 sec, 0 rows)

SL burned out? (AP results when SL is ON) (0 sec, 26 rows)

SerialYYYYJJJSL CurrentSL Voltage
2232020052-12.1524.4
2232020053-12.1424.43
2232020054-12.1424.43
2232020055-12.1424.4
2232020056-12.1524.4
2232020057-12.1524.4
2232020058-12.1524.43
2232020059-12.1124.43

Non-critical warnings from b-files' comments for days between 2020059 and 2020059 (1 sec, 2 rows)

TypeCommentcountpriorityfnamelocationserialtimevalueminmax
rl:2020 058 06 11 30 Reset Warm -> Tepid from 2020 058 06 11 24 until 2020 058 06 11 25 . Counts: 1,013B05920.223EUREK#22306:06:151-1
cfg_out:Illegal command (cfg199B05920.223EUREK#22306:06:391-1

Measurements and tests summary.


_jday_loc_serial_skcaodedsdtedfzhghprssluvzs
059EUREKA PEARL#223ecsc442123201927415
(1 sec)

Relative and absolute humidity for day 2020059 (2 sec, 1 rows)

SerialRH, %AH, g/m37-day RH historyDes-nt replcd days ago
2235.10.5