Database last updated on Fri, 11 Sep 2026 00:11:05 +0000 by localhost

Critical warnings from b-files' comments for days between 2019313 and 2019313 (0 sec, 2 rows)

TypeCommentcountpriorityfnamelocationserialtimevalueminmax
rl:2019 312 05 42 52 Motor #2: appears stuck with a reference sensor activated.21B31319.223EUREK#22305:31:461-1
rl:2019 312 06 12 57 PMT counter failure11B31319.223EUREK#22305:31:581-1

Instrument constant changes for the week before 2019313 (2 sec, 0 rows)

SL burned out? (AP results when SL is ON) (0 sec, 32 rows)

SerialYYYYJJJSL CurrentSL Voltage
2232019306-12.1424.4
2232019307-12.1424.43
2232019308-12.1724.4
2232019309-12.1724.4
2232019310-12.1524.4
2232019311-12.1524.4
2232019312-12.1424.43
2232019313-12.1524.4

Non-critical warnings from b-files' comments for days between 2019313 and 2019313 (0 sec, 4 rows)

TypeCommentcountpriorityfnamelocationserialtimevalueminmax
rl:2019 312 05 41 43 Reset Warm -> Tepid from 2019 312 05 41 38 until 2019 312 05 41 38 . Counts: 1,013B31319.223EUREK#22305:31:441-1
rl:2019 312 06 07 27 New value set for watermark LOOP.RUN.TIME: 653.999996 millisec.43B31319.223EUREK#22305:31:501-1
rl:2019 312 06 07 27 New value set for watermark LOOP.RUN.TIME: 653.999996 millisec.43B31319.223EUREK#22305:31:501-1
cfg_out:Illegal command (cfg199B31319.223EUREK#22305:32:051-1

Measurements and tests summary.


_jday_loc_serial_skcdtedhghprssl
313EUREKA PEARL#223ecsc218826
(1 sec)

New operators? (0 sec, 88 rows)

Relative and absolute humidity for day 2019313 (1 sec, 1 rows)

SerialRH, %AH, g/m37-day RH historyDes-nt replcd days ago
2235.70.8