Database last updated on Thu, 10 Sep 2026 23:11:10 +0000 by localhost

Critical warnings from b-files' comments for days between 2019301 and 2019301 (0 sec, 2 rows)

TypeCommentcountpriorityfnamelocationserialtimevalueminmax
rl:2019 300 05 39 41 Motor #2: appears stuck with a reference sensor activated.31B30119.223EUREK#22305:35:561-1
rl:2019 301 17 25 51 PMT counter failure11B30119.223EUREK#22305:36:091-1

Instrument constant changes for the week before 2019301 (3 sec, 0 rows)

SL burned out? (AP results when SL is ON) (0 sec, 32 rows)

SerialYYYYJJJSL CurrentSL Voltage
2232019294-12.1724.4
2232019295-12.1524.4
2232019296-12.1524.43
2232019297-12.1524.4
2232019298-12.1524.4
2232019299-12.1524.43
2232019300-12.1524.4
2232019301-12.1424.43

Non-critical warnings from b-files' comments for days between 2019301 and 2019301 (0 sec, 4 rows)

TypeCommentcountpriorityfnamelocationserialtimevalueminmax
rl:2019 300 05 38 33 Reset Warm -> Tepid from 2019 300 05 38 27 until 2019 300 05 38 28 . Counts: 1,013B30119.223EUREK#22305:35:541-1
rl:2019 300 06 06 54 New value set for watermark LOOP.RUN.TIME: 700.000000 millisec.33B30119.223EUREK#22305:36:011-1
rl:2019 300 06 06 54 New value set for watermark LOOP.RUN.TIME: 700.000000 millisec.33B30119.223EUREK#22305:36:011-1
cfg_out:Illegal command (cfg199B30119.223EUREK#22305:36:161-1

Measurements and tests summary.


_jday_loc_serial_skcaodedsdtedhghprssluv
301EUREKA PEARL#223ecsc2221982516
(0 sec)

Relative and absolute humidity for day 2019301 (2 sec, 1 rows)

SerialRH, %AH, g/m37-day RH historyDes-nt replcd days ago
2236.20.9