Critical warnings from b-files' comments for days between 2019205 and 2019205 (0 sec, 4 rows)

TypeCommentcountpriorityfnamelocationserialtimevalueminmax
rl:2019 204 05 58 14 Motor #2: appears stuck with a reference sensor activated.21B20519.223EUREK#22305:56:461-1
rl:2019 204 06 15 05 New value set for watermark LOOP.RUN.TIME: 674.000025 millisec.73B20519.223EUREK#22305:56:501-1
rl:2019 204 06 15 05 New value set for watermark LOOP.RUN.TIME: 674.000025 millisec.73B20519.223EUREK#22305:56:501-1
rl:2019 204 06 55 31 PMT counter failure11B20519.223EUREK#22305:57:041-1

Instrument constant changes for the week before 2019205 (1 sec, 0 rows)

SL burned out? (AP results when SL is ON) (0 sec, 32 rows)

SerialYYYYJJJSL CurrentSL Voltage
2232019198-12.1724.4
2232019199-12.1724.4
2232019200-12.1724.37
2232019201-12.1824.37
2232019202-12.1724.4
2232019203-12.1524.4
2232019204-12.1524.4
2232019205-12.1724.37

Non-critical warnings from b-files' comments for days between 2019205 and 2019205 (0 sec, 5 rows)

TypeCommentcountpriorityfnamelocationserialtimevalueminmax
rl:2019 204 05 57 05 Reset Warm -> Tepid from 2019 204 05 56 59 until 2019 204 05 57 00 . Counts: 1,013B20519.223EUREK#22305:56:431-1
rl:2019 205 09 39 58 New value set for watermark LOOP.RUN.TIME: 716.999960 millisec.13B20519.223EUREK#22305:57:061-1
rl:2019 205 09 39 58 New value set for watermark LOOP.RUN.TIME: 716.999960 millisec.13B20519.223EUREK#22305:57:061-1
cfg_out:Illegal command (cfg199B20519.223EUREK#22305:57:121-1
ds:DS intensity too high ( 370587/cy) quitting.199B20519.223EUREK#22316:22:241-1

Measurements and tests summary.


_jday_loc_serial_skcaodedsdtedhghprssluvzs
205EUREKA PEARL#223ecsc1021031166157452
(0 sec)

Relative and absolute humidity for day 2019205 (2 sec, 1 rows)

SerialRH, %AH, g/m37-day RH historyDes-nt replcd days ago
2236.21.2