Database last updated on Mon, 14 Sep 2026 14:41:01 +0000 by localhost
Critical warnings from b-files' comments for days between 2019010 and 2019010 (0 sec, 4 rows)
| Type | Comment | count | priority | fname | location | serial | time | value | min | max |
| rl: | 2019 009 05 44 24 Motor #2: appears stuck with a reference sensor activated. | 2 | 1 | B01019.223 | TORON | #223 | 05:35:19 | | 1 | -1 |
| rl: | 2019 009 06 01 15 New value set for watermark LOOP.RUN.TIME: 699.000025 millisec. | 8 | 3 | B01019.223 | TORON | #223 | 05:35:23 | | 1 | -1 |
| rl: | 2019 009 06 01 15 New value set for watermark LOOP.RUN.TIME: 699.000025 millisec. | 8 | 3 | B01019.223 | TORON | #223 | 05:35:23 | | 1 | -1 |
| rl: | 2019 009 06 51 05 PMT counter failure | 1 | 1 | B01019.223 | TORON | #223 | 05:35:32 | | 1 | -1 |
Instrument constant changes for the week before 2019010 (4 sec, 0 rows)
SL burned out? (AP results when SL is ON) (0 sec, 8 rows)
| Serial | YYYYJJJ | SL Current | SL Voltage |
| 223 | 2019003 | -12.2 | 24.34 |
| 223 | 2019004 | -12.17 | 24.37 |
| 223 | 2019005 | -12.18 | 24.4 |
| 223 | 2019006 | -12.18 | 24.37 |
| 223 | 2019007 | -12.17 | 24.4 |
| 223 | 2019008 | -12.18 | 24.37 |
| 223 | 2019009 | -12.17 | 24.37 |
| 223 | 2019010 | -12.17 | 24.4 |
HG burned out? (AP results when HG is ON) (0 sec, 21 rows)
Hg with step corrections larger than 10 steps or multiple corrections larger than 2.5 steps (0 sec, 1 rows)
| Serial | YYYYJJJ | HG large | max correction |
| 223 | 2019010 | 4 | 16.5 |
Non-critical warnings from b-files' comments for days between 2019010 and 2019010 (0 sec, 6 rows)
| Type | Comment | count | priority | fname | location | serial | time | value | min | max |
| rl: | 2019 009 05 43 14 Reset Warm -> Tepid from 2019 009 05 43 08 until 2019 009 05 43 09 . Counts: 1,0 | 1 | 3 | B01019.223 | TORON | #223 | 05:35:17 | | 1 | -1 |
| rl: | 2019 010 15 08 59 New value set for watermark LOOP.RUN.TIME: 724.000025 millisec. | 2 | 3 | B01019.223 | TORON | #223 | 05:35:38 | | 1 | -1 |
| rl: | 2019 010 15 08 59 New value set for watermark LOOP.RUN.TIME: 724.000025 millisec. | 2 | 3 | B01019.223 | TORON | #223 | 05:35:38 | | 1 | -1 |
| cfg_out: | Illegal command (cfg | 1 | 99 | B01019.223 | TORON | #223 | 05:35:47 | | 1 | -1 |
| zs: | ZS intensity too low ( 253/cy), skipping. | 4 | 99 | B01019.223 | TORON | #223 | 13:10:00 | | 1 | -1 |
| ds: | Suppressed oscillation of filter wheel. | 2 | 5 | B01019.223 | TORON | #223 | 18:59:37 | | 1 | -1 |
Measurements and tests summary.
| _jday | _loc | _serial | _skc | ds | dt | ed | fz | hg | hp | rs | sl | uv | zs |
| 010 | TORONTO | #223 | ecsc | 31 | 2 | 1 | 1 | 15 | 12 | 2 | 6 | 40 | 18 |
(1 sec)
New operators? (0 sec, 103 rows)
Relative and absolute humidity for day 2019010 (2 sec, 1 rows)
| Serial | RH, % | AH, g/m3 | 7-day RH history | Des-nt replcd days ago |
| 223 | 12.8 | 2.0 |