Database last updated on Mon, 14 Sep 2026 05:11:04 +0000 by localhost

Critical warnings from b-files' comments for days between 2018156 and 2018156 (0 sec, 2 rows)

TypeCommentcountpriorityfnamelocationserialtimevalueminmax
sc:SC intensity too low ( 9/cy), skipping.63B15618.223Delft#22306:12:301-1
rl:2018 156 00 19 46 Motor #2: appears stuck with a reference sensor activated.21B15618.223Delft#22323:41:261-1

Instrument constant changes for the week before 2018156 (5 sec, 0 rows)

SL burned out? (AP results when SL is ON) (0 sec, 3 rows)

Hg with step corrections larger than 10 steps or multiple corrections larger than 2.5 steps (0 sec, 1 rows)

SerialYYYYJJJHG largemax correction
223201815627.3

Non-critical warnings from b-files' comments for days between 2018156 and 2018156 (0 sec, 4 rows)

TypeCommentcountpriorityfnamelocationserialtimevalueminmax
ds:DS intensity too high ( 292539/cy) quitting.399B15618.223Delft#22311:31:341-1
ds:Suppressed oscillation of filter wheel.15B15618.223Delft#22313:27:201-1
rl:2018 156 00 18 10 MOTOR.MAX.JERK: Symbol not found.23B15618.223Delft#22323:41:241-1
chk_cfg:MOTOR.MAX.JERK[10]=(symbol not found)23B15618.223Delft#22323:41:361-1

Measurements and tests summary.


_jday_loc_serial_skcdsdtedhghprssluv
156Delft#223dsmk3ec34111363610499
(0 sec)

New operators? (0 sec, 56 rows)

Relative and absolute humidity for day 2018156 (1 sec, 1 rows)

SerialRH, %AH, g/m37-day RH historyDes-nt replcd days ago
2237.51.9