Database last updated on Mon, 14 Sep 2026 03:41:10 +0000 by localhost
Critical warnings from b-files' comments for days between 2018155 and 2018155 (0 sec, 2 rows)
| Type | Comment | count | priority | fname | location | serial | time | value | min | max |
| sc: | SC intensity too low ( 5/cy), skipping. | 6 | 3 | B15518.223 | Delft | #223 | 06:12:52 | | 1 | -1 |
| rl: | 2018 155 00 19 27 Motor #2: appears stuck with a reference sensor activated. | 2 | 1 | B15518.223 | Delft | #223 | 23:41:19 | | 1 | -1 |
Instrument constant changes for the week before 2018155 (1 sec, 0 rows)
SL burned out? (AP results when SL is ON) (0 sec, 4 rows)
Hg with step corrections larger than 10 steps or multiple corrections larger than 2.5 steps (0 sec, 1 rows)
| Serial | YYYYJJJ | HG large | max correction |
| 223 | 2018155 | 2 | 7.3 |
Non-critical warnings from b-files' comments for days between 2018155 and 2018155 (0 sec, 2 rows)
| Type | Comment | count | priority | fname | location | serial | time | value | min | max |
| rl: | 2018 155 00 18 07 MOTOR.MAX.JERK: Symbol not found. | 2 | 3 | B15518.223 | Delft | #223 | 23:41:17 | | 1 | -1 |
| chk_cfg: | MOTOR.MAX.JERK[10]=(symbol not found) | 2 | 3 | B15518.223 | Delft | #223 | 23:41:29 | | 1 | -1 |
Measurements and tests summary.
| _jday | _loc | _serial | _skc | ds | dt | ed | hg | hp | rs | sl | uv |
| 155 | Delft | #223 | dsmk3ec | 35 | 10 | 1 | 37 | 35 | 10 | 49 | 9 |
(0 sec)
New operators? (0 sec, 56 rows)
Relative and absolute humidity for day 2018155 (2 sec, 1 rows)
| Serial | RH, % | AH, g/m3 | 7-day RH history | Des-nt replcd days ago |
| 223 | 7.5 | 2.2 |