Database last updated on Mon, 14 Sep 2026 16:11:05 +0000 by localhost

Critical warnings from b-files' comments for days between 2018137 and 2018137 (0 sec, 3 rows)

TypeCommentcountpriorityfnamelocationserialtimevalueminmax
rl:2018 137 13 09 27 Motor #3: appears stuck with a reference sensor activated.31B13718.223Delft#22323:39:161-1
rl:2018 137 13 24 30 New value set for watermark LOOP.RUN.TIME: 653.999996 millisec.73B13718.223Delft#22323:39:201-1
rl:2018 137 13 24 30 New value set for watermark LOOP.RUN.TIME: 653.999996 millisec.73B13718.223Delft#22323:39:201-1

Instrument constant changes for the week before 2018137 (3 sec, 0 rows)

SL burned out? (AP results when SL is ON) (0 sec, 4 rows)

Hg with step corrections larger than 10 steps or multiple corrections larger than 2.5 steps (0 sec, 1 rows)

SerialYYYYJJJHG largemax correction
2232018137121.8

Non-critical warnings from b-files' comments for days between 2018137 and 2018137 (0 sec, 6 rows)

TypeCommentcountpriorityfnamelocationserialtimevalueminmax
sl:DEL key pressed199B13718.223Delft#22313:13:331-1
ds:DS intensity too high ( 403720/cy) quitting.299B13718.223Delft#22315:56:291-1
rl:2018 137 13 09 26 Reset Tepid -> Tepid from 2018 137 13 09 11 until 2018 137 13 09 21 . Counts: 1,013B13718.223Delft#22323:39:151-1
rl:2018 137 13 09 27 Motor #3: appears stuck with a reference sensor activated.13B13718.223Delft#22323:39:161-1
rl:2018 137 13 09 27 Motor #11: appears stuck with a reference sensor activated.13B13718.223Delft#22323:39:181-1
chk_cfg:MOTOR.MAX.JERK[10]=(symbol not found)23B13718.223Delft#22323:39:351-1

Measurements and tests summary.


_jday_loc_serial_skcdsdtedhghprssluv
137Delft#223dsmk3ec2,Starting schedule dsmk3ec2164114144205
(0 sec)

New operators? (0 sec, 82 rows)

Relative and absolute humidity for day 2018137 (2 sec, 1 rows)

SerialRH, %AH, g/m37-day RH historyDes-nt replcd days ago
22333.55.8