Database last updated on Mon, 14 Sep 2026 07:41:08 +0000 by localhost

Critical warnings from b-files' comments for days between 2018131 and 2018131 (0 sec, 2 rows)

TypeCommentcountpriorityfnamelocationserialtimevalueminmax
sl:Lamp not on... sl test terminated F1= 0 F4= 10031B13118.223Delft#22310:08:401-1
rl:2018 131 00 24 13 Motor #2: appears stuck with a reference sensor activated.21B13118.223Delft#22323:39:151-1

Instrument constant changes for the week before 2018131 (2 sec, 0 rows)

SL burned out? (AP results when SL is ON) (0 sec, 6 rows)

Hg with step corrections larger than 10 steps or multiple corrections larger than 2.5 steps (0 sec, 1 rows)

SerialYYYYJJJHG largemax correction
223201813128.9

HP with step corrections larger than 20 steps (0 sec, 1 rows)

SerialYYYYJJJHP largemax correction
22320181311380.0

Non-critical warnings from b-files' comments for days between 2018131 and 2018131 (0 sec, 7 rows)

TypeCommentcountpriorityfnamelocationserialtimevalueminmax
td:Time correction 3.7 seconds13B13118.223Delft#22300:29:593.7-33
ls:Illegal command (ls) in schedule dsmk3ec, ignoring199B13118.223Delft#22306:54:041-1
ds:DS intensity too high ( 243987/cy) quitting.299B13118.223Delft#22313:19:541-1
sc:SC intensity too low ( 410/cy), skipping.23B13118.223Delft#22314:48:351-1
ds:Suppressed oscillation of filter wheel.15B13118.223Delft#22315:04:141-1
rl:2018 131 00 22 39 MOTOR.MAX.JERK: Symbol not found.23B13118.223Delft#22323:39:131-1
chk_cfg:MOTOR.MAX.JERK[10]=(symbol not found)23B13118.223Delft#22323:39:251-1

Measurements and tests summary.


_jday_loc_serial_skcdsdtedhghprssluv
131Delft#223dsmk3ec39111335110419
(0 sec)

New operators? (0 sec, 143 rows)

Relative and absolute humidity for day 2018131 (1 sec, 1 rows)

SerialRH, %AH, g/m37-day RH historyDes-nt replcd days ago
22327.38.9