Database last updated on Mon, 14 Sep 2026 08:40:56 +0000 by localhost

Critical warnings from b-files' comments for days between 2018130 and 2018130 (0 sec, 2 rows)

TypeCommentcountpriorityfnamelocationserialtimevalueminmax
sc:SC intensity too low ( 7/cy), skipping.53B13018.223Delft#22306:34:581-1
rl:2018 130 00 24 14 Motor #2: appears stuck with a reference sensor activated.21B13018.223Delft#22323:39:171-1

Instrument constant changes for the week before 2018130 (3 sec, 0 rows)

SL burned out? (AP results when SL is ON) (0 sec, 7 rows)

Hg with step corrections larger than 10 steps or multiple corrections larger than 2.5 steps (0 sec, 1 rows)

SerialYYYYJJJHG largemax correction
223201813027.0

HP with step corrections larger than 20 steps (0 sec, 1 rows)

SerialYYYYJJJHP largemax correction
2232018130445.9

Non-critical warnings from b-files' comments for days between 2018130 and 2018130 (0 sec, 3 rows)

TypeCommentcountpriorityfnamelocationserialtimevalueminmax
ds:DS intensity too high ( 238741/cy) quitting.399B13018.223Delft#22309:40:311-1
rl:2018 130 00 23 00 MOTOR.MAX.JERK: Symbol not found.23B13018.223Delft#22323:39:151-1
chk_cfg:MOTOR.MAX.JERK[10]=(symbol not found)23B13018.223Delft#22323:39:261-1

Measurements and tests summary.


_jday_loc_serial_skcdsdtedhghprssluv
130Delft#223dsmk3ec34101333910449
(0 sec)

New operators? (0 sec, 130 rows)

Relative and absolute humidity for day 2018130 (1 sec, 1 rows)

SerialRH, %AH, g/m37-day RH historyDes-nt replcd days ago
22327.26.0