Database last updated on Mon, 14 Sep 2026 21:11:07 +0000 by localhost

Critical warnings from b-files' comments for days between 2018124 and 2018124 (0 sec, 3 rows)

TypeCommentcountpriorityfnamelocationserialtimevalueminmax
sl:Lamp not on... sl test terminated F1= 0 F4= 1511B12418.223Delft#22310:28:151-1
hp:Spectrometer alignment trouble.11B12418.223Delft#22310:29:411-1
rl:2018 124 00 24 04 Motor #2: appears stuck with a reference sensor activated.31B12418.223Delft#22323:39:361-1

Instrument constant changes for the week before 2018124 (3 sec, 0 rows)

SL burned out? (AP results when SL is ON) (0 sec, 7 rows)

HG burned out? (AP results when HG is ON) (0 sec, 9 rows)

Hg with step corrections larger than 10 steps or multiple corrections larger than 2.5 steps (0 sec, 1 rows)

SerialYYYYJJJHG largemax correction
2232018124310.1

HP with step corrections larger than 20 steps (0 sec, 1 rows)

SerialYYYYJJJHP largemax correction
22320181241070.3

Non-critical warnings from b-files' comments for days between 2018124 and 2018124 (0 sec, 5 rows)

TypeCommentcountpriorityfnamelocationserialtimevalueminmax
ls:Illegal command (ls) in schedule dsmk3ec, ignoring199B12418.223Delft#22307:04:081-1
fr:Micrometer #2 discrepancy= -724 Now set to offset = 0199B12418.223Delft#22310:33:28-724-1010
rl:2018 124 00 22 47 MOTOR.MAX.JERK: Symbol not found.23B12418.223Delft#22323:39:341-1
rl:2018 124 10 42 24 Motor #9: appears stuck with a reference sensor activated.13B12418.223Delft#22323:39:391-1
chk_cfg:MOTOR.MAX.JERK[10]=(symbol not found)23B12418.223Delft#22323:39:471-1

Measurements and tests summary.


_jday_loc_serial_skcdsdtedhghprssluv
124Delft#223dsmk3ec541013646104010
(0 sec)

New operators? (0 sec, 72 rows)

Relative and absolute humidity for day 2018124 (2 sec, 1 rows)

SerialRH, %AH, g/m37-day RH historyDes-nt replcd days ago
22327.17.9