Critical warnings from b-files' comments for days between 2018123 and 2018123 (0 sec, 4 rows)

TypeCommentcountpriorityfnamelocationserialtimevalueminmax
sl:Lamp not on... sl test terminated F1= 0 F4= 3941B12318.223Delft#22310:29:351-1
hp:Spectrometer alignment trouble.11B12318.223Delft#22310:35:291-1
18:Date changed from 03/05/18 to 02/05/1811B12318.223Delft#22314:48:291-1
rl:2018 123 00 22 53 Motor #2: appears stuck with a reference sensor activated.21B12318.223Delft#22323:39:411-1

Instrument constant changes for the week before 2018123 (1 sec, 0 rows)

SL burned out? (AP results when SL is ON) (0 sec, 7 rows)

HG burned out? (AP results when HG is ON) (0 sec, 10 rows)

Hg with step corrections larger than 10 steps or multiple corrections larger than 2.5 steps (0 sec, 1 rows)

SerialYYYYJJJHG largemax correction
223201812336.4

HP with step corrections larger than 20 steps (0 sec, 1 rows)

SerialYYYYJJJHP largemax correction
22320181232380.0

Non-critical warnings from b-files' comments for days between 2018123 and 2018123 (0 sec, 5 rows)

TypeCommentcountpriorityfnamelocationserialtimevalueminmax
fr:Micrometer #2 discrepancy= -37 Now set to offset = 0199B12318.223Delft#22301:09:53-37-1010
ls:Illegal command (ls) in schedule dsmk3ec, ignoring199B12318.223Delft#22307:05:351-1
18:Time correction 41827.7 seconds13B12318.223Delft#22314:48:2941827.7-33
rl:2018 123 00 21 40 MOTOR.MAX.JERK: Symbol not found.23B12318.223Delft#22323:39:391-1
chk_cfg:MOTOR.MAX.JERK[10]=(symbol not found)23B12318.223Delft#22323:39:511-1

Measurements and tests summary.


_jday_loc_serial_skcdsdtedhghprssluv
123Delft#223menu531013358103710
(0 sec)

New operators? (0 sec, 70 rows)

Relative and absolute humidity for day 2018123 (2 sec, 1 rows)

SerialRH, %AH, g/m37-day RH historyDes-nt replcd days ago
22326.86.3