Database last updated on Mon, 14 Sep 2026 08:40:56 +0000 by localhost

Critical warnings from b-files' comments for days between 2018118 and 2018118 (0 sec, 2 rows)

TypeCommentcountpriorityfnamelocationserialtimevalueminmax
sc:SC intensity too low ( 16/cy), skipping.63B11818.223Delft#22306:53:271-1
rl:2018 118 00 24 06 Motor #2: appears stuck with a reference sensor activated.21B11818.223Delft#22323:40:181-1

Instrument constant changes for the week before 2018118 (3 sec, 0 rows)

SL burned out? (AP results when SL is ON) (0 sec, 3 rows)

HG burned out? (AP results when HG is ON) (0 sec, 7 rows)

Hg with step corrections larger than 10 steps or multiple corrections larger than 2.5 steps (0 sec, 1 rows)

SerialYYYYJJJHG largemax correction
223201811825.8

HP with step corrections larger than 20 steps (0 sec, 1 rows)

SerialYYYYJJJHP largemax correction
2232018118120.4

Non-critical warnings from b-files' comments for days between 2018118 and 2018118 (0 sec, 3 rows)

TypeCommentcountpriorityfnamelocationserialtimevalueminmax
ls:Illegal command (ls) in schedule dsmk3ec, ignoring499B11818.223Delft#22306:53:431-1
rl:2018 118 00 22 53 MOTOR.MAX.JERK: Symbol not found.23B11818.223Delft#22323:40:161-1
chk_cfg:MOTOR.MAX.JERK[10]=(symbol not found)23B11818.223Delft#22323:40:281-1

Measurements and tests summary.


_jday_loc_serial_skcdsdtedhghprssluv
118Delft#223dsmk3ec111113638114910
(1 sec)

New operators? (0 sec, 12 rows)

Relative and absolute humidity for day 2018118 (1 sec, 1 rows)

SerialRH, %AH, g/m37-day RH historyDes-nt replcd days ago
22326.53.7