Database last updated on Tue, 15 Sep 2026 00:43:04 +0000 by localhost

Critical warnings from b-files' comments for days between 2018117 and 2018117 (1 sec, 1 rows)

TypeCommentcountpriorityfnamelocationserialtimevalueminmax
rl:2018 117 00 24 23 Motor #2: appears stuck with a reference sensor activated.21B11718.223Delft#22323:40:251-1

Instrument constant changes for the week before 2018117 (3 sec, 0 rows)

SL burned out? (AP results when SL is ON) (0 sec, 2 rows)

HG burned out? (AP results when HG is ON) (0 sec, 6 rows)

Hg with step corrections larger than 10 steps or multiple corrections larger than 2.5 steps (0 sec, 1 rows)

SerialYYYYJJJHG largemax correction
223201811727.7

HP with step corrections larger than 20 steps (0 sec, 1 rows)

SerialYYYYJJJHP largemax correction
2232018117566.1

Non-critical warnings from b-files' comments for days between 2018117 and 2018117 (0 sec, 6 rows)

TypeCommentcountpriorityfnamelocationserialtimevalueminmax
td:Time correction 6.4 seconds13B11718.223Delft#22300:30:136.4-33
ls:Illegal command (ls) in schedule dsmk3ec, ignoring199B11718.223Delft#22307:15:341-1
sc:Suppressed oscillation of filter wheel.15B11718.223Delft#22313:53:491-1
sc:SC intensity too low ( 345/cy), skipping.33B11718.223Delft#22314:54:441-1
rl:2018 117 00 22 48 MOTOR.MAX.JERK: Symbol not found.23B11718.223Delft#22323:40:231-1
chk_cfg:MOTOR.MAX.JERK[10]=(symbol not found)23B11718.223Delft#22323:40:341-1

Measurements and tests summary.


_jday_loc_serial_skcdsdtedhghprssluv
117Delft#223dsmk3ec361113540104610
(1 sec)

New operators? (0 sec, 14 rows)

Relative and absolute humidity for day 2018117 (3 sec, 1 rows)

SerialRH, %AH, g/m37-day RH historyDes-nt replcd days ago
22326.04.8