Database last updated on Mon, 14 Sep 2026 13:41:00 +0000 by localhost

Critical warnings from b-files' comments for days between 2018116 and 2018116 (0 sec, 3 rows)

TypeCommentcountpriorityfnamelocationserialtimevalueminmax
sl:Lamp not on... sl test terminated F1= 0 F4= 16021B11618.223Delft#22310:59:361-1
hp:Spectrometer alignment trouble.11B11618.223Delft#22312:07:131-1
rl:2018 116 00 20 23 Motor #2: appears stuck with a reference sensor activated.21B11618.223Delft#22323:40:351-1

Instrument constant changes for the week before 2018116 (4 sec, 0 rows)

SL burned out? (AP results when SL is ON) (0 sec, 2 rows)

HG burned out? (AP results when HG is ON) (0 sec, 7 rows)

Hg with step corrections larger than 10 steps or multiple corrections larger than 2.5 steps (0 sec, 1 rows)

SerialYYYYJJJHG largemax correction
223201811638.4

HP with step corrections larger than 20 steps (0 sec, 1 rows)

SerialYYYYJJJHP largemax correction
2232018116280.0

Non-critical warnings from b-files' comments for days between 2018116 and 2018116 (0 sec, 6 rows)

TypeCommentcountpriorityfnamelocationserialtimevalueminmax
ls:Illegal command (ls) in schedule dsmk3ec, ignoring299B11618.223Delft#22307:12:131-1
sc:SC intensity too low ( 431/cy), skipping.23B11618.223Delft#22308:41:081-1
ds:Suppressed oscillation of filter wheel.15B11618.223Delft#22310:01:471-1
fr:Micrometer #2 discrepancy= -279 Now set to offset = 0199B11618.223Delft#22312:11:04-279-1010
rl:2018 116 00 19 17 MOTOR.MAX.JERK: Symbol not found.23B11618.223Delft#22323:40:331-1
chk_cfg:MOTOR.MAX.JERK[10]=(symbol not found)23B11618.223Delft#22323:40:441-1

Measurements and tests summary.


_jday_loc_serial_skcdsdtedhghprssluv
116Delft#223dsmk3ec4491353894210
(0 sec)

New operators? (0 sec, 14 rows)

Relative and absolute humidity for day 2018116 (2 sec, 1 rows)

SerialRH, %AH, g/m37-day RH historyDes-nt replcd days ago
22325.84.2