Database last updated on Mon, 14 Sep 2026 16:11:05 +0000 by localhost
Critical warnings from b-files' comments for days between 2018111 and 2018111 (0 sec, 1 rows)
| Type | Comment | count | priority | fname | location | serial | time | value | min | max |
| rl: | 2018 111 00 17 41 Motor #2: appears stuck with a reference sensor activated. | 2 | 1 | B11118.223 | Delft | #223 | 23:52:08 | | 1 | -1 |
Instrument constant changes for the week before 2018111 (2 sec, 0 rows)
HG burned out? (AP results when HG is ON) (0 sec, 9 rows)
Hg with step corrections larger than 10 steps or multiple corrections larger than 2.5 steps (0 sec, 1 rows)
| Serial | YYYYJJJ | HG large | max correction |
| 223 | 2018111 | 30 | 7.2 |
Non-critical warnings from b-files' comments for days between 2018111 and 2018111 (0 sec, 4 rows)
| Type | Comment | count | priority | fname | location | serial | time | value | min | max |
| ls: | Illegal command (ls) in schedule dsmk3ec, ignoring | 1 | 99 | B11118.223 | Delft | #223 | 07:26:52 | | 1 | -1 |
| fr: | Micrometer #2 discrepancy= 27 Now set to offset = 0 | 1 | 99 | B11118.223 | Delft | #223 | 18:53:41 | 27 | -10 | 10 |
| rl: | 2018 111 00 16 35 MOTOR.MAX.JERK: Symbol not found. | 2 | 3 | B11118.223 | Delft | #223 | 23:52:06 | | 1 | -1 |
| chk_cfg: | MOTOR.MAX.JERK[10]=(symbol not found) | 2 | 3 | B11118.223 | Delft | #223 | 23:52:18 | | 1 | -1 |
Measurements and tests summary.
| _jday | _loc | _serial | _skc | ds | dt | ed | hg | hp | rs | sl | uv |
| 111 | Delft | #223 | dsmk3ec | 63 | 8 | 1 | 86 | 61 | 8 | 32 | 7 |
(1 sec)
New operators? (0 sec, 28 rows)
Relative and absolute humidity for day 2018111 (1 sec, 1 rows)
| Serial | RH, % | AH, g/m3 | 7-day RH history | Des-nt replcd days ago |
| 223 | 25.6 | 8.0 |