Database last updated on Mon, 14 Sep 2026 22:11:13 +0000 by localhost

Critical warnings from b-files' comments for days between 2018024 and 2018024 (1 sec, 6 rows)

TypeCommentcountpriorityfnamelocationserialtimevalueminmax
hg:Lamp not detected f1= 0 f4= 3121B02418.223Delft#22306:18:531-1
hg:Lamp not on... hg test terminated F1= 0 F4= 3131B02418.223Delft#22306:18:531-1
sl:SL lamp is dead: the current is 011B02418.223Delft#22321:58:581-1
rl:2018 024 00 11 42 Motor #2: appears stuck with a reference sensor activated.31B02418.223Delft#22323:55:031-1
rl:2018 024 00 30 14 New value set for watermark LOOP.RUN.TIME: 699.000025 millisec.103B02418.223Delft#22323:55:051-1
rl:2018 024 00 30 14 New value set for watermark LOOP.RUN.TIME: 699.000025 millisec.103B02418.223Delft#22323:55:051-1

Instrument constant changes for the week before 2018024 (4 sec, 0 rows)

Hg with step corrections larger than 10 steps or multiple corrections larger than 2.5 steps (0 sec, 1 rows)

SerialYYYYJJJHG largemax correction
2232018024928.0

HP with step corrections larger than 20 steps (0 sec, 1 rows)

SerialYYYYJJJHP largemax correction
2232018024660.3

Non-critical warnings from b-files' comments for days between 2018024 and 2018024 (1 sec, 3 rows)

TypeCommentcountpriorityfnamelocationserialtimevalueminmax
td:Time correction 3.4 seconds13B02418.223Delft#22302:10:083.4-33
rl:2018 024 00 10 49 Reset Warm -> Tepid from 2018 024 00 10 43 until 2018 024 00 10 43 . Counts: 1,013B02418.223Delft#22323:55:021-1
chk_cfg:MOTOR.MAX.JERK[10]=(symbol not found)23B02418.223Delft#22323:55:261-1

Measurements and tests summary.


_jday_loc_serial_skcdtedhghprssluv
024Delft#223dsmk3ec9131309336
(2 sec)

New operators? (0 sec, 12 rows)

Relative and absolute humidity for day 2018024 (3 sec, 1 rows)

SerialRH, %AH, g/m37-day RH historyDes-nt replcd days ago
22321.72.9