Database last updated on Mon, 14 Sep 2026 19:41:03 +0000 by localhost
Critical warnings from b-files' comments for days between 2017356 and 2017356 (0 sec, 3 rows)
| Type | Comment | count | priority | fname | location | serial | time | value | min | max |
| sl: | Lamp not on... sl test terminated F1= 0 F4= 0 | 1 | 1 | B35617.223 | Delft | #223 | 21:14:44 | | 1 | -1 |
| sl: | SL lamp is dead: the current is 0 | 1 | 1 | B35617.223 | Delft | #223 | 21:14:44 | | 1 | -1 |
| rl: | 2017 355 23 57 34 Motor #2: appears stuck with a reference sensor activated. | 3 | 1 | B35617.223 | Delft | #223 | 23:41:48 | | 1 | -1 |
Instrument constant changes for the week before 2017356 (3 sec, 0 rows)
Hg with step corrections larger than 10 steps or multiple corrections larger than 2.5 steps (0 sec, 1 rows)
| Serial | YYYYJJJ | HG large | max correction |
| 223 | 2017356 | 4 | 4.1 |
Non-critical warnings from b-files' comments for days between 2017356 and 2017356 (0 sec, 4 rows)
| Type | Comment | count | priority | fname | location | serial | time | value | min | max |
| rl: | 2017 355 23 56 44 Reset Warm -> Tepid from 2017 355 23 56 38 until 2017 355 23 56 39 . Counts: 1,0 | 1 | 3 | B35617.223 | Delft | #223 | 23:41:47 | | 1 | -1 |
| rl: | 2017 356 00 27 47 New value set for watermark LOOP.RUN.TIME: 700.000000 millisec. | 4 | 3 | B35617.223 | Delft | #223 | 23:41:51 | | 1 | -1 |
| rl: | 2017 356 00 27 47 New value set for watermark LOOP.RUN.TIME: 700.000000 millisec. | 4 | 3 | B35617.223 | Delft | #223 | 23:41:51 | | 1 | -1 |
| chk_cfg: | MOTOR.MAX.JERK[10]=(symbol not found) | 2 | 3 | B35617.223 | Delft | #223 | 23:42:04 | | 1 | -1 |
Measurements and tests summary.
| _jday | _loc | _serial | _skc | dt | ed | hg | hp | rs | sl | uv |
| 356 | Delft | #223 | dsmk3ec | 8 | 1 | 27 | 22 | 8 | 37 | 5 |
(0 sec)
New operators? (0 sec, 14 rows)
Relative and absolute humidity for day 2017356 (1 sec, 1 rows)
| Serial | RH, % | AH, g/m3 | 7-day RH history | Des-nt replcd days ago |
| 223 | 12.0 | 1.5 |