Database last updated on Sun, 08 Feb 2026 02:10:54 +0000 by localhost

Critical warnings from b-files' comments for days between 2021292 and 2021292 (0 sec, 1 rows)

TypeCommentcountpriorityfnamelocationserialtimevalueminmax
rl:2021 292 01 00 16 Motor #3: appears stuck with a reference sensor activated.31B29221.206SATUR#20608:13:061-1

Instrument constant changes for the week before 2021292 (3 sec, 0 rows)

SL burned out? (AP results when SL is ON) (0 sec, 3 rows)

SerialYYYYJJJSL CurrentSL Voltage
2062021290012.99
20620212911.4910.69
20620212921.4910.67

Non-critical warnings from b-files' comments for days between 2021292 and 2021292 (0 sec, 11 rows)

TypeCommentcountpriorityfnamelocationserialtimevalueminmax
rl:2021 292 01 00 15 Reset No -> Tepid from 2021 292 00 59 27 until 2042 255 18 31 54 . Counts: 1,013B29221.206SATUR#20608:13:011-1
rl:2021 292 01 00 16 Checksum bad in low level command.13B29221.206SATUR#20608:13:031-1
rl:2021 292 01 00 16 Motor #3: appears stuck with a reference sensor activated.13B29221.206SATUR#20608:13:061-1
rl:2021 292 01 00 16 Motor #11: appears stuck with a reference sensor activated.13B29221.206SATUR#20608:13:081-1
rl:2021 292 01 00 30 Reset Warm -> Warm from 2021 292 01 00 27 until 2021 292 01 00 28 . Counts: 1,133B29221.206SATUR#20608:13:111-1
rl:2021 292 01 00 45 Motor #9: appears stuck with a reference sensor activated.13B29221.206SATUR#20608:13:161-1
zs:ZS intensity too low ( 174/cy), skipping.399B29221.206SATUR#20614:41:281-1
ux:HOME key pressed199B29221.206SATUR#20615:39:331-1
ux:UV extended scan aborted13B29221.206SATUR#20615:40:141-1
ux:Going to menu199B29221.206SATUR#20615:40:581-1
si:New SI, HC = 12, NC = -2940 Old = -2 , -2955199B29221.206SATUR#20615:52:251-1

Measurements and tests summary.


_jday_loc_serial_skcdsdtedfmfzhghprssluvzs
292SATURNA#206ecsc,Next schedule will be ecsc (now ecsc ZS= 93 ZS= 93.318)23123712115164516
(0 sec)

Relative and absolute humidity for day 2021292 (1 sec, 1 rows)

SerialRH, %AH, g/m37-day RH historyDes-nt replcd days ago
20645.94.0