Database last updated on Wed, 26 Aug 2026 00:11:00 +0000 by localhost

Critical warnings from b-files' comments for days between 2026206 and 2026206 (0 sec, 3 rows)

TypeCommentcountpriorityfnamelocationserialtimevalueminmax
rl:2026 206 17 12 50 Motor #3: appears stuck with a reference sensor activated.21B20626.205RESOL#20506:37:001-1
fz:Suppressed oscillation of filter wheel.55B20626.205RESOL#20510:52:171-1
sl:Brewer failed to respond 5 times11B20626.205RESOL#20513:13:261-1

Instrument constant changes for the week before 2026206 (3 sec, 0 rows)

SL burned out? (AP results when SL is ON) (0 sec, 2 rows)

Hg with step corrections larger than 10 steps or multiple corrections larger than 2.5 steps (0 sec, 1 rows)

SerialYYYYJJJHG largemax correction
205202620635.8

Non-critical warnings from b-files' comments for days between 2026206 and 2026206 (0 sec, 4 rows)

TypeCommentcountpriorityfnamelocationserialtimevalueminmax
rl:2026 206 17 12 50 Reset No -> Tepid from 2026 206 13 10 31 until 2057 170 10 44 28 . Counts: 1,013B20626.205RESOL#20506:36:551-1
rl:2026 206 17 12 50 Checksum bad in low level command.13B20626.205RESOL#20506:36:571-1
rl:2026 206 17 12 50 Motor #3: appears stuck with a reference sensor activated.13B20626.205RESOL#20506:37:001-1
rl:2026 206 17 12 51 Motor #11: appears stuck with a reference sensor activated.13B20626.205RESOL#20506:37:021-1

Measurements and tests summary.


_jday_loc_serial_skcaodedsedfzhghpsluvzs
206RESOLUTE#205ecsc108126232151363651
(0 sec)

Relative and absolute humidity for day 2026206 (1 sec, 1 rows)

SerialRH, %AH, g/m37-day RH historyDes-nt replcd days ago
2054.60.7