Database last updated on Mon, 26 Jan 2026 12:10:52 +0000 by localhost

Critical warnings from b-files' comments for days between 2025194 and 2025194 (0 sec, 9 rows)

TypeCommentcountpriorityfnamelocationserialtimevalueminmax
rl:2025 193 21 20 35 New value set for watermark SUPPLY.VOLTAGE.HIGH53B19425.205RESOL#20500:21:331-1
hp:Spectrometer alignment trouble.11B19425.205RESOL#20500:31:501-1
hg:HG failed81B19425.205RESOL#20500:37:19failed1-1
rl:2025 192 18 34 20 Motor #3: appears stuck with a reference sensor activated.81B19425.205RESOL#20512:42:251-1
rl:2025 192 20 06 25 Attempt to exceed min boundary143B19425.205RESOL#20512:42:351-1
rl:2025 192 20 06 25 <M,10,-1736>153B19425.205RESOL#20512:42:371-1
rl:2025 192 20 06 25 <M,10,-1736>145B19425.205RESOL#20512:42:371-1
hv: Lamp not on... hv test terminated F1= 5396 F4= 469871B19425.205RESOL#20520:38:411-1
ew:Command file ew not found --- continuing11B19425.205RESOL#20520:57:081-1

Instrument constant changes for the week before 2025194 (5 sec, 0 rows)

Hg with step corrections larger than 10 steps or multiple corrections larger than 2.5 steps (0 sec, 1 rows)

SerialYYYYJJJHG largemax correction
205202519414018.1

HP with step corrections larger than 20 steps (0 sec, 1 rows)

SerialYYYYJJJHP largemax correction
2052025194280.0

Non-critical warnings from b-files' comments for days between 2025194 and 2025194 (0 sec, 20 rows)

TypeCommentcountpriorityfnamelocationserialtimevalueminmax
rl:2025 193 21 17 39 Reset No -> Tepid from 2025 193 21 17 25 until 2016 193 21 17 34 . Counts: 1,013B19425.205RESOL#20500:21:161-1
rl:2025 193 21 17 39 Checksum bad in low level command.13B19425.205RESOL#20500:21:181-1
rl:2025 193 21 17 40 Motor #3: appears stuck with a reference sensor activated.13B19425.205RESOL#20500:21:201-1
rl:2025 193 21 17 40 Motor #11: appears stuck with a reference sensor activated.13B19425.205RESOL#20500:21:221-1
rl:2025 193 21 18 25 Reset Warm -> Warm from 2025 193 21 18 22 until 2025 193 21 18 22 . Counts: 1,113B19425.205RESOL#20500:21:261-1
rl:2025 193 21 18 26 Motor #10: appears stuck with a reference sensor activated.13B19425.205RESOL#20500:21:281-1
rl:2025 193 21 20 35 MAIN.C : motor initialization timed out.15B19425.205RESOL#20500:21:301-1
rl:2025 192 18 34 20 Reset No -> Tepid from 2025 192 18 15 16 until 2016 192 18 34 14 . Counts: 1,013B19425.205RESOL#20512:42:211-1
rl:2025 192 18 34 20 Checksum bad in low level command.13B19425.205RESOL#20512:42:231-1
rl:2025 192 18 34 20 Motor #3: appears stuck with a reference sensor activated.13B19425.205RESOL#20512:42:251-1
rl:2025 192 18 34 21 Motor #11: appears stuck with a reference sensor activated.13B19425.205RESOL#20512:42:281-1
rl:2025 192 18 34 56 Reset Warm -> Warm from 2025 192 18 34 53 until 2025 192 18 34 54 . Counts: 1,113B19425.205RESOL#20512:42:311-1
rl:2025 192 18 37 06 MAIN.C : motor initialization timed out.15B19425.205RESOL#20512:42:331-1
rl:2025 192 22 32 43 Motor #10: appears stuck with a reference sensor activated.23B19425.205RESOL#20512:43:061-1
rl:2025 192 22 33 33 Motor #9: appears stuck with a reference sensor activated.13B19425.205RESOL#20512:43:101-1
rl:2025 193 00 30 45 Attempt to exceed min boundary43B19425.205RESOL#20512:43:261-1
rl:2025 193 00 30 45 <M,10,-1736>45B19425.205RESOL#20512:43:281-1
rl:2025 193 12 48 57 Bad command string. Command string buffer cleared.13B19425.205RESOL#20514:25:431-1
rl:2025 193 12 48 57 Bad command string. Command string buffer cleared.13B19425.205RESOL#20514:25:431-1
ew:Command file ew not found --- continuing199B19425.205RESOL#20520:57:08continuing1-1

Measurements and tests summary.


_jday_loc_serial_skcdthghprssl
194RESOLUTE#205menu3531032
(1 sec)

Relative and absolute humidity for day 2025194 (1 sec, 1 rows)

SerialRH, %AH, g/m37-day RH historyDes-nt replcd days ago
20525.84.7