Database last updated on Wed, 04 Feb 2026 13:40:51 +0000 by localhost

Critical warnings from b-files' comments for days between 2023171 and 2023171 (0 sec, 1 rows)

TypeCommentcountpriorityfnamelocationserialtimevalueminmax
rl:2023 171 01 15 04 Motor #3: appears stuck with a reference sensor activated.21B17123.205RESOL#20506:29:421-1

Instrument constant changes for the week before 2023171 (3 sec, 0 rows)

SL burned out? (AP results when SL is ON) (0 sec, 2 rows)

HG burned out? (AP results when HG is ON) (0 sec, 7 rows)

Hg with step corrections larger than 10 steps or multiple corrections larger than 2.5 steps (0 sec, 1 rows)

SerialYYYYJJJHG largemax correction
205202317166.7

Non-critical warnings from b-files' comments for days between 2023171 and 2023171 (0 sec, 10 rows)

TypeCommentcountpriorityfnamelocationserialtimevalueminmax
zs:ZS measurement terminated15B17123.205RESOL#20501:11:401-1
rl:2023 171 01 15 04 Checksum bad in low level command.13B17123.205RESOL#20506:29:401-1
rl:2023 171 01 15 04 Motor #3: appears stuck with a reference sensor activated.13B17123.205RESOL#20506:29:421-1
rl:2023 171 01 15 04 Motor #11: appears stuck with a reference sensor activated.13B17123.205RESOL#20506:29:451-1
pd:Command file pd not found199B17123.205RESOL#20513:16:35found1-1
hg:HOME key pressed399B17123.205RESOL#20519:31:211-1
pd:Going to menu399B17123.205RESOL#20519:31:241-1
si:New SI, HC =199B17123.205RESOL#20519:57:361-1
ds:DS intensity too high ( 277010/cy) quitting.299B17123.205RESOL#20520:40:531-1
ds:Suppressed oscillation of filter wheel.15B17123.205RESOL#20523:37:211-1

Measurements and tests summary.


_jday_loc_serial_skcaodedsdtedhghprsslzs
171RESOLUTE#205menu,Starting schedule calds40422227222564
(0 sec)

Relative and absolute humidity for day 2023171 (1 sec, 1 rows)

SerialRH, %AH, g/m37-day RH historyDes-nt replcd days ago
2051.80.3