Database last updated on Thu, 10 Sep 2026 22:10:49 +0000 by localhost
Critical warnings from b-files' comments for days between 2019299 and 2019299 (0 sec, 1 rows)
| Type | Comment | count | priority | fname | location | serial | time | value | min | max |
| hp: | Spectrometer alignment trouble. | 1 | 1 | B29919.205 | RESOL | #205 | 21:40:35 | | 1 | -1 |
Instrument constant changes for the week before 2019299 (2 sec, 0 rows)
SL burned out? (AP results when SL is ON) (0 sec, 24 rows)
Hg with step corrections larger than 10 steps or multiple corrections larger than 2.5 steps (0 sec, 1 rows)
| Serial | YYYYJJJ | HG large | max correction |
| 205 | 2019299 | 2 | 3.5 |
Non-critical warnings from b-files' comments for days between 2019299 and 2019299 (0 sec, 4 rows)
| Type | Comment | count | priority | fname | location | serial | time | value | min | max |
| rl: | 2019 298 06 11 47 Reset Warm -> Tepid from 2019 298 06 11 41 until 2019 298 06 11 42 . Counts: 1,0 | 1 | 3 | B29919.205 | RESOL | #205 | 06:10:55 | | 1 | -1 |
| zs: | ZS intensity too low ( 253/cy), skipping. | 3 | 99 | B29919.205 | RESOL | #205 | 15:10:02 | | 1 | -1 |
| ux: | HOME key pressed | 1 | 99 | B29919.205 | RESOL | #205 | 21:22:36 | | 1 | -1 |
| ux: | Going to menu | 1 | 99 | B29919.205 | RESOL | #205 | 21:23:29 | | 1 | -1 |
Measurements and tests summary.
| _jday | _loc | _serial | _skc | aode | ds | dt | ed | fz | hg | hp | rs | sl | uv | zs |
| 299 | RESOLUTE | #205 | ecsc,Starting schedule ecsc | 11 | 11 | 2 | 1 | 21 | 9 | 7 | 2 | 6 | 30 | 11 |
(1 sec)
Relative and absolute humidity for day 2019299 (1 sec, 1 rows)
| Serial | RH, % | AH, g/m3 | 7-day RH history | Des-nt replcd days ago |
| 205 | 9.2 | 0.8 |