Critical warnings from b-files' comments for days between 2023316 and 2023316 (0 sec, 4 rows)

TypeCommentcountpriorityfnamelocationserialtimevalueminmax
rl:2023 315 04 07 20 Motor #2: appears stuck with a reference sensor activated.11#20403:55:381-1
rl:2023 315 04 30 31 New value set for watermark LOOP.RUN.TIME: 661.999989 millisec.63#20403:55:411-1
rl:2023 315 04 30 31 New value set for watermark LOOP.RUN.TIME: 661.999989 millisec.63#20403:55:411-1
rl:2023 316 05 29 33 PMT counter failure11#20403:55:541-1

Instrument constant changes for the week before 2023316 (1 sec, 0 rows)

HG burned out? (AP results when HG is ON) (0 sec, 6 rows)

Non-critical warnings from b-files' comments for days between 2023316 and 2023316 (0 sec, 5 rows)

TypeCommentcountpriorityfnamelocationserialtimevalueminmax
rl:2023 315 04 06 04 Bad command string. Command string buffer cleared.13#20403:55:341-1
rl:2023 315 04 06 04 Bad command string. Command string buffer cleared.13#20403:55:341-1
rl:2023 315 04 06 04 <]13#20403:55:351-1
rl:2023 316 05 32 46 New value set for watermark LOOP.RUN.TIME: 716.999960 millisec.13#20403:55:561-1
rl:2023 316 05 32 46 New value set for watermark LOOP.RUN.TIME: 716.999960 millisec.13#20403:55:561-1

Measurements and tests summary.


_jday_loc_serial_skcaodedsdtedfzhghprssluvzs
316GOOSE BAY#204ecsc,Next schedule will be ecsc (now ecsc ZS= 93 ZS= 93.32)15152221514274316
(0 sec)

Dead time (1 sec)

Nothing to report.

Standard lamp R6 (1 sec)

Nothing to report.

Standard lamp R5 (1 sec)

Nothing to report.

Number of days since last dome/window cleaning (0 sec, 2 rows)

serialdays
204more than a year

Missing files (0 sec, 1 rows)

Seriallast datadays ago
2042023314002

Run/Stop results for last week. (0 sec, 1 rows)

SerialTotalFailed 0Failed darkFailed 1Failed 2Failed 3Failed 4Failed 5Failed double
2041600000000

AP results for SB Brewers. (0 sec, 1 rows)

SerialJdayt1 offt1 ont2 offt2 ont3 offt3 onHV offHV on+12V off+12V on+5V off+5V on-12V off-12V on+24V off+24V on+5V sps off+5V sps onSL A offSL A onSL V offSL V onWatts
204202331617.317.318.418.413.713.71490.41490.412.212.25.15.1-12.0-12.024.224.25.05.00.01.50.010.5016.0

Standard lamp intensity analysis (0 sec, 1 rows)

Serialminmaxfilterfixable30-day history
#204289328028971200yes