Critical warnings from b-files' comments for days between 2026077 and 2026077 (0 sec, 4 rows)

TypeCommentcountpriorityfnamelocationserialtimevalueminmax
hp:Spectrometer alignment trouble.11B07726.203CHURC#20306:49:231-1
re:Previous RE didn't seem to help, so not doing reset again. Operator intervention required. Quitting to menu.11B07726.203CHURC#20306:51:401-1
rl:2026 076 06 38 42 Motor #10: appears stuck with a reference sensor activated.141B07726.203CHURC#20312:31:241-1
rl:2026 076 06 38 42 Motor #10: appears stuck with a reference sensor activated.143B07726.203CHURC#20312:31:241-1

Instrument constant changes for the week before 2026077 (6 sec, 0 rows)

Non-critical warnings from b-files' comments for days between 2026077 and 2026077 (0 sec, 5 rows)

TypeCommentcountpriorityfnamelocationserialtimevalueminmax
pd:Going to menu299B07726.203CHURC#20306:51:411-1
hp:HOME key pressed199B07726.203CHURC#20311:58:171-1
rl:2026 076 06 37 50 Reset Warm -> Tepid from 2026 076 06 37 44 until 2026 076 06 37 44 . Counts: 1,033B07726.203CHURC#20312:31:211-1
ds:Suppressed oscillation of filter wheel.35B07726.203CHURC#20315:26:171-1
ds:DS intensity too low ( 59/cy), skipping.499B07726.203CHURC#20323:34:511-1

Measurements and tests summary.


_jday_loc_serial_skcaodedsedfzhghpsluvzs
077CHURCHILL#203ecsc,Next schedule will be ecsc (now ecsc ZS= 93 ZS= 93.03)7882213141252542
(1 sec)

Relative and absolute humidity for day 2026077 (2 sec, 1 rows)

SerialRH, %AH, g/m37-day RH historyDes-nt replcd days ago
20313.32.0