Critical warnings from b-files' comments for days between 2023238 and 2023238 (0 sec, 2 rows)
| Type | Comment | count | priority | fname | location | serial | time | value | min | max |
| rl: | 2023 238 17 20 29 Motor #11: appears stuck with a reference sensor activated. | 2 | 1 | | | #203 | 06:25:16 | | 1 | -1 |
| rl: | 2023 238 02 04 09 New value set for watermark SUPPLY.VOLTAGE.HIGH[ | 22 | 3 | | | #203 | 06:25:29 | | 1 | -1 |
Instrument constant changes for the week before 2023238 (2 sec, 0 rows)
HG burned out? (AP results when HG is ON) (0 sec, 29 rows)
Hg with step corrections larger than 10 steps or multiple corrections larger than 2.5 steps (0 sec, 1 rows)
| Serial | YYYYJJJ | HG large | max correction |
| 203 | 2023238 | 4 | 7.1 |
Non-critical warnings from b-files' comments for days between 2023238 and 2023238 (0 sec, 12 rows)
| Type | Comment | count | priority | fname | location | serial | time | value | min | max |
| rl: | 2023 238 17 20 29 Motor #11: appears stuck with a reference sensor activated. | 1 | 3 | | | #203 | 06:25:16 | | 1 | -1 |
| rl: | 2023 238 17 22 37 MAIN.C : motor initialization timed out. | 1 | 5 | | | #203 | 06:25:18 | | 1 | -1 |
| dp: | Doing dispersion with internal HG. | 1 | 99 | | | #203 | 09:38:12 | | 1 | -1 |
| skc: | DEL key pressed | 1 | 99 | | | #203 | 10:39:18 | | 1 | -1 |
| skc: | HOME key pressed | 3 | 99 | | | #203 | 12:07:35 | | 1 | -1 |
| dsp: | Doing dispersion test with Hg (UV port) | 2 | 99 | | | #203 | 12:08:01 | | 1 | -1 |
| checkhg: | Last HG had to adjust the micrometer by at least 1 | 1 | 99 | | | #203 | 14:17:58 | | 1 | -1 |
| pd: | Command file pd not found | 1 | 99 | | | #203 | 17:25:37 | found | 1 | -1 |
| si: | New SI, HC = 0, NC = 1475 Old = 2 , 1480 | 1 | 99 | | | #203 | 17:44:27 | | 1 | -1 |
| pd: | Going to menu | 2 | 99 | | | #203 | 17:44:52 | | 1 | -1 |
| ds: | Suppressed oscillation of filter wheel. | 1 | 5 | | | #203 | 18:17:35 | | 1 | -1 |
| zs: | ZS measurement terminated | 1 | 5 | | | #203 | 18:53:52 | | 1 | -1 |
Measurements and tests summary.
| _jday | _loc | _serial | _skc | aode | ds | dt | ed | fz | hg | hp | rs | sl | uv | zs |
| 238 | CHURCHILL | #203 | 42 | 42 | 2 | 2 | 2 | 35 | 33 | 2 | 4 | 1 | 30 |
(1 sec)
Dead time (0 sec)
Nothing to report.
Standard lamp R6 (1 sec)
Nothing to report.
Standard lamp R5 (1 sec)
Nothing to report.
Number of days since last dome/window cleaning (0 sec, 3 rows)
| serial | days |
| 203 | more than a year |
| 3 | more than a year |
Missing files (0 sec, 1 rows)
| Serial | last data | days ago |
| 203 | 2023237 | 001 |
Questionable Run/Stop results. (0 sec, 1 rows)
| Serial | Jday | slit 0 | dark rs | dark counts | slit 1 | slit 2 | slit 3 | slit 4 | slit 5 | slits 3+5 |
| 203 | 2023238 | 0.994 | 1.062 | 81 / 86 (1.8M) | 0.999 | 1.000 | 1.000 | 1.000 | 0.999 | 1.000 |
Run/Stop results for last week. (0 sec, 1 rows)
| Serial | Total | Failed 0 | Failed dark | Failed 1 | Failed 2 | Failed 3 | Failed 4 | Failed 5 | Failed double |
| 203 | 27 | 0 | 1 | 0 | 0 | 0 | 0 | 0 | 0 |
AP results for SB Brewers. (0 sec, 1 rows)
| Serial | Jday | t1 off | t1 on | t2 off | t2 on | t3 off | t3 on | HV off | HV on | +12V off | +12V on | +5V off | +5V on | -12V off | -12V on | +24V off | +24V on | +5V sps off | +5V sps on | SL A off | SL A on | SL V off | SL V on | Watts |
| 203 | 2023238 | 27.6 | 27.6 | 27.3 | 27.2 | 26.4 | 26.4 | 1452.0 | 1452.5 | 12.1 | 12.1 | 5.1 | 5.1 | -11.7 | -11.7 | 24.1 | 24.1 | 5.0 | 5.0 | 0.0 | 1.4 | 0.0 | 9.16 | 13.2 |
Standard lamp intensity analysis (0 sec, 1 rows)
| Serial | min | max | filter | fixable | 30-day history |
| #203 | 2826510 | 2828590 | 0 | yes |