Database last updated on Tue, 03 Feb 2026 17:10:51 +0000 by localhost

Critical warnings from b-files' comments for days between 2023030 and 2023030 (0 sec, 6 rows)

TypeCommentcountpriorityfnamelocationserialtimevalueminmax
hp:Spectrometer alignment trouble.71B03023.203CHURC#20302:15:101-1
rl:2023 030 14 13 44 Attempt to exceed min boundary213B03023.203CHURC#20306:41:271-1
rl:2023 030 14 13 44 <M,10,213B03023.203CHURC#20306:41:291-1
rl:2023 030 14 13 44 <M,10,215B03023.203CHURC#20306:41:291-1
hg:Lamp not detected f1= 2 f4= 9131B03023.203CHURC#20315:58:491-1
hg:Lamp not on... hg test terminated F1= 2 F4= 9311B03023.203CHURC#20315:58:491-1

Instrument constant changes for the week before 2023030 (7 sec, 0 rows)

HP with step corrections larger than 20 steps (0 sec, 1 rows)

SerialYYYYJJJHP largemax correction
20320230306779.9

Non-critical warnings from b-files' comments for days between 2023030 and 2023030 (0 sec, 1 rows)

TypeCommentcountpriorityfnamelocationserialtimevalueminmax
rl:2023 030 10 28 58 New value set for watermark SUPPLY.VOLTAGE.HIGH[13B03023.203CHURC#20306:41:251-1

Measurements and tests summary.


_jday_loc_serial_skcdtedfmfzhghprssluv
030CHURCHILL#203ecsc,Must do HG now,Next schedule will be ecsc (now ecsc ZS= 93 ZS= 94.012)1232241081358
(1 sec)

Relative and absolute humidity for day 2023030 (2 sec, 1 rows)

SerialRH, %AH, g/m37-day RH historyDes-nt replcd days ago
2035.50.5