Database last updated on Wed, 26 Aug 2026 13:11:02 +0000 by localhost

Critical warnings from b-files' comments for days between 2026157 and 2026157 (0 sec, 10 rows)

TypeCommentcountpriorityfnamelocationserialtimevalueminmax
hp:Spectrometer alignment trouble.31B15726.195STONY#19512:15:261-1
rl:2022 156 23 19 15 Motor #3: appears stuck with a reference sensor activated.41B15726.195STONY#19512:21:441-1
rl:2022 156 23 25 38 Bad command string. Command string buffer cleared.163B15726.195STONY#19512:21:501-1
rl:2022 156 23 25 38 Bad command string. Command string buffer cleared.163B15726.195STONY#19512:21:501-1
rl:2022 156 23 25 38 <G,1056>283B15726.195STONY#19512:21:511-1
rl:2022 156 23 40 43 Attempt to exceed min boundary123B15726.195STONY#19512:22:411-1
rl:2022 156 23 40 43 <M,9,-1740>125B15726.195STONY#19512:22:431-1
hg:No intensity detected. Check PMT.11B15726.195STONY#19515:40:301-1
hg:Lamp not on... hg test terminated F1= 0 F4= 021B15726.195STONY#19515:40:301-1
hg:Lamp not detected f1= 0 f4= 021B15726.195STONY#19515:40:311-1

Instrument constant changes for the week before 2026157 (3 sec, 0 rows)

SL burned out? (AP results when SL is ON) (0 sec, 5 rows)

HP with step corrections larger than 20 steps (0 sec, 1 rows)

SerialYYYYJJJHP largemax correction
1952026157450.0

Non-critical warnings from b-files' comments for days between 2026157 and 2026157 (0 sec, 15 rows)

TypeCommentcountpriorityfnamelocationserialtimevalueminmax
rl:2022 157 22 48 32 Reset No -> Tepid from 2022 157 22 45 19 until 2028 157 22 48 27 . Counts: 1,013B15726.195STONY#19507:46:011-1
rl:2022 157 22 48 32 Checksum bad in low level command.13B15726.195STONY#19507:46:031-1
rl:2022 157 22 48 32 Motor #3: appears stuck with a reference sensor activated.13B15726.195STONY#19507:46:061-1
rl:2022 157 22 48 33 Motor #11: appears stuck with a reference sensor activated.13B15726.195STONY#19507:46:091-1
hg:HOME key pressed299B15726.195STONY#19512:17:151-1
pd:Going to menu299B15726.195STONY#19512:17:191-1
rl:2022 156 23 19 15 Reset No -> Tepid from 2022 156 23 00 25 until 2028 156 23 19 10 . Counts: 1,013B15726.195STONY#19512:21:401-1
rl:2022 156 23 19 15 Checksum bad in low level command.13B15726.195STONY#19512:21:421-1
rl:2022 156 23 19 15 Motor #3: appears stuck with a reference sensor activated.13B15726.195STONY#19512:21:441-1
rl:2022 156 23 19 16 Motor #11: appears stuck with a reference sensor activated.13B15726.195STONY#19512:21:461-1
rl:2022 156 23 22 50 Reset Warm -> Warm from 2022 156 23 22 47 until 2022 156 23 22 47 . Counts: 0,113B15726.195STONY#19512:21:481-1
rl:2022 156 23 40 43 Attempt to exceed min boundary13B15726.195STONY#19512:22:411-1
rl:2022 156 23 40 43 <M,9,-1740>15B15726.195STONY#19512:22:431-1
dsp:Doing dispersion test with Cd (UV port)299B15726.195STONY#19516:06:431-1
rl:2022 157 10 51 30 Reset Warm -> Tepid from 2022 157 10 51 24 until 2022 157 10 51 25 . Counts: 1,023B15726.195STONY#19519:56:351-1

Measurements and tests summary.


_jday_loc_serial_skcdtedhghprssl
157STONY PLAIN#195menu52282856
(0 sec)

Relative and absolute humidity for day 2026157 (2 sec, 1 rows)

SerialRH, %AH, g/m37-day RH historyDes-nt replcd days ago
19523.44.8