Database last updated on Fri, 30 Jan 2026 18:10:53 +0000 by localhost

Critical warnings from b-files' comments for days between 2024248 and 2024248 (0 sec, 4 rows)

TypeCommentcountpriorityfnamelocationserialtimevalueminmax
rl:2024 248 15 15 30 Motor #3: appears stuck with a reference sensor activated.21B24824.194ALERT#19404:14:551-1
rl:2024 248 15 16 34 New value set for watermark SUPPLY.VOLTAGE.HIGH[53B24824.194ALERT#19404:15:001-1
re:Reset failed 5 times. Quitting and going to the menu.11B24824.194ALERT#19414:10:461-1
pd:Brewer failed to respond 5 times21B24824.194ALERT#19414:12:591-1

Instrument constant changes for the week before 2024248 (4 sec, 0 rows)

Hg with step corrections larger than 10 steps or multiple corrections larger than 2.5 steps (0 sec, 1 rows)

SerialYYYYJJJHG largemax correction
194202424825.6

Non-critical warnings from b-files' comments for days between 2024248 and 2024248 (1 sec, 4 rows)

TypeCommentcountpriorityfnamelocationserialtimevalueminmax
rl:2024 248 15 15 30 Checksum bad in low level command.13B24824.194ALERT#19404:14:531-1
rl:2024 248 15 15 30 Motor #3: appears stuck with a reference sensor activated.13B24824.194ALERT#19404:14:551-1
rl:2024 248 15 15 31 Motor #11: appears stuck with a reference sensor activated.13B24824.194ALERT#19404:14:571-1
ab_sk:HOME key pressed199B24824.194ALERT#19415:30:561-1

Measurements and tests summary.


_jday_loc_serial_skcaodedsedfzhghpsluvzs
248ALERT#194ecsc3838242151142627
(1 sec)

Relative and absolute humidity for day 2024248 (1 sec, 1 rows)

SerialRH, %AH, g/m37-day RH historyDes-nt replcd days ago
19410.61.0