Database last updated on Sat, 31 Jan 2026 12:10:51 +0000 by localhost

Critical warnings from b-files' comments for days between 2024163 and 2024163 (0 sec, 1 rows)

TypeCommentcountpriorityfnamelocationserialtimevalueminmax
rl:2024 163 10 28 50 Motor #3: appears stuck with a reference sensor activated.21#19404:15:571-1

Instrument constant changes for the week before 2024163 (3 sec, 0 rows)

Hg with step corrections larger than 10 steps or multiple corrections larger than 2.5 steps (0 sec, 1 rows)

SerialYYYYJJJHG largemax correction
194202416323.6

Non-critical warnings from b-files' comments for days between 2024163 and 2024163 (0 sec, 8 rows)

TypeCommentcountpriorityfnamelocationserialtimevalueminmax
rl:2024 163 10 28 49 Checksum bad in low level command.13#19404:15:541-1
rl:2024 163 10 28 50 Motor #3: appears stuck with a reference sensor activated.13#19404:15:571-1
rl:2024 163 10 28 50 Motor #11: appears stuck with a reference sensor activated.13#19404:15:591-1
ds:HOME key pressed499#19408:38:231-1
pd:Going to menu499#19408:38:281-1
si:New SI, HC = 71, NC =499#19408:50:221-1
ds:Suppressed oscillation of filter wheel.45#19409:03:321-1
ds:DS intensity too high ( 279128/cy) quitting.299#19419:38:131-1

Measurements and tests summary.


_jday_loc_serial_skcaodedsdtedhghprssluvzs
163ALERT#194calo3152154122218152126
(0 sec)

Relative and absolute humidity for day 2024163 (2 sec, 1 rows)

SerialRH, %AH, g/m37-day RH historyDes-nt replcd days ago
19423.42.1