Database last updated on Sun, 01 Feb 2026 03:40:51 +0000 by localhost

Critical warnings from b-files' comments for days between 2024032 and 2024032 (0 sec, 4 rows)

TypeCommentcountpriorityfnamelocationserialtimevalueminmax
hp:Spectrometer alignment trouble.81#19416:51:301-1
hg:Lamp not on... hg test terminated F1= 1 F4= 221#19416:54:111-1
hg:Lamp not detected f1= 1 f4= 221#19416:54:121-1
re:Previous RE didn't seem to help, so not doing reset again. Operator intervention required. Quitting to menu.21#19417:11:371-1

Instrument constant changes for the week before 2024032 (2 sec, 0 rows)

HG burned out? (AP results when HG is ON) (0 sec, 18 rows)

SerialYYYYJJJHG CurrentHG Voltage
19420240250.1813.81
19420240260.1813.81
19420240270.1813.81
19420240280.1813.8
19420240290.1813.81
19420240300.1813.81
19420240310.1813.8
19420240310.3113.81
19420240310.1813.81
19420240310.1813.78
19420240320.1813.81
19420240320.1813.8

Non-critical warnings from b-files' comments for days between 2024032 and 2024032 (0 sec, 2 rows)

TypeCommentcountpriorityfnamelocationserialtimevalueminmax
zs:ZS intensity too low ( 104/cy), skipping.199#19412:36:121-1
pd:Going to menu399#19416:38:581-1

Measurements and tests summary.


_jday_loc_serial_skcdtfmfzhghprssluvzs
032Toronto#194Attempting recovery FR,ecsc,Going to menu,Must do HG now217266231615
(0 sec)

Relative and absolute humidity for day 2024032 (2 sec, 1 rows)

SerialRH, %AH, g/m37-day RH historyDes-nt replcd days ago
1945.70.9