Database last updated on Sun, 01 Feb 2026 07:40:55 +0000 by localhost
Critical warnings from b-files' comments for days between 2024011 and 2024011 (0 sec, 3 rows)
| Type | Comment | count | priority | fname | location | serial | time | value | min | max |
| hg: | Lamp not on... hg test terminated F1= 0 F4= 0 | 1 | 1 | | | #194 | 10:38:45 | | 1 | -1 |
| hg: | No intensity detected. Check PMT. | 1 | 1 | | | #194 | 10:38:46 | | 1 | -1 |
| hg: | Lamp not detected f1= 0 f4= 0 | 1 | 1 | | | #194 | 10:38:47 | | 1 | -1 |
Instrument constant changes for the week before 2024011 (3 sec, 0 rows)
SL burned out? (AP results when SL is ON) (0 sec, 7 rows)
HG burned out? (AP results when HG is ON) (0 sec, 10 rows)
| Serial | YYYYJJJ | HG Current | HG Voltage |
| 194 | 2024004 | 0.18 | 13.84 |
| 194 | 2024005 | 0.18 | 13.84 |
| 194 | 2024006 | 0.18 | 13.84 |
| 194 | 2024007 | 0.18 | 13.86 |
| 194 | 2024008 | 0.18 | 13.83 |
| 194 | 2024009 | 0.18 | 13.83 |
| 194 | 2024010 | 0.18 | 13.83 |
| 194 | 2024010 | 0.18 | 13.84 |
| 194 | 2024011 | 0.18 | 13.83 |
| 194 | 2024011 | 0.18 | 13.84 |
Non-critical warnings from b-files' comments for days between 2024011 and 2024011 (0 sec, 3 rows)
| Type | Comment | count | priority | fname | location | serial | time | value | min | max |
| rl: | 2024 010 05 43 23 Bad command string. Command string buffer cleared. | 1 | 3 | | | #194 | 05:35:11 | | 1 | -1 |
| rl: | 2024 010 05 43 23 Bad command string. Command string buffer cleared. | 1 | 3 | | | #194 | 05:35:11 | | 1 | -1 |
| rl: | 2024 010 05 43 23 < | 1 | 3 | | | #194 | 05:35:13 | | 1 | -1 |
Measurements and tests summary.
| _jday | _loc | _serial | _skc | dt | ed | fz | hg | hp | rs | sl | uv | zs |
| 011 | Toronto | #194 | ecsc,Next schedule will be ecsc (now ecsc ZS= 93 ZS= 93.033) | 2 | 2 | 2 | 18 | 18 | 2 | 5 | 43 | 36 |
(0 sec)
Relative and absolute humidity for day 2024011 (2 sec, 1 rows)
| Serial | RH, % | AH, g/m3 | 7-day RH history | Des-nt replcd days ago |
| 194 | 15.8 | 1.5 |