Database last updated on Mon, 02 Feb 2026 02:40:49 +0000 by localhost

Critical warnings from b-files' comments for days between 2023198 and 2023198 (0 sec, 3 rows)

TypeCommentcountpriorityfnamelocationserialtimevalueminmax
rl:2023 198 13 59 49 Motor #3: appears stuck with a reference sensor activated.21#19405:31:001-1
rl:2023 198 16 40 31 PMT counter failure11#19405:31:071-1
ds:Suppressed oscillation of filter wheel.75#19414:51:571-1

Instrument constant changes for the week before 2023198 (2 sec, 0 rows)

HG burned out? (AP results when HG is ON) (0 sec, 20 rows)

SerialYYYYJJJHG CurrentHG Voltage
19420231910.1813.77
19420231920.1813.75
19420231930.1913.78
19420231940.1813.8
19420231950.1813.78
19420231970.1813.78
19420231970.3213.75
19420231970.3213.74
19420231970.3313.75
19420231970.3313.78
19420231970.3313.77
19420231980.1813.78
19420231980.1913.77
19420231980.1913.78
19420231980.3213.78
19420231980.1913.74
19420231980.3213.8
19420231980.3313.75
19420231980.3313.78
19420231980.3313.77

HP with step corrections larger than 20 steps (0 sec, 1 rows)

SerialYYYYJJJHP largemax correction
1942023198139.1

Non-critical warnings from b-files' comments for days between 2023198 and 2023198 (0 sec, 4 rows)

TypeCommentcountpriorityfnamelocationserialtimevalueminmax
rl:2023 198 13 59 48 Checksum bad in low level command.13#19405:30:571-1
rl:2023 198 13 59 49 Motor #3: appears stuck with a reference sensor activated.13#19405:31:001-1
rl:2023 198 13 59 49 Motor #11: appears stuck with a reference sensor activated.13#19405:31:031-1
rl:2023 198 14 00 53 New value set for watermark SUPPLY.VOLTAGE.HIGH[13#19405:31:051-1

Measurements and tests summary.


_jday_loc_serial_skcaodedsdtedfzhghprsslzs
198Toronto#194calds,Next schedule will be ecds (now calds ZS= 93 ZS= 93.041)12913922125252735
(0 sec)

Relative and absolute humidity for day 2023198 (1 sec, 1 rows)

SerialRH, %AH, g/m37-day RH historyDes-nt replcd days ago
19427.97.6