Database last updated on Thu, 10 Sep 2026 21:40:59 +0000 by localhost

Critical warnings from b-files' comments for days between 2019186 and 2019186 (0 sec, 1 rows)

TypeCommentcountpriorityfnamelocationserialtimevalueminmax
rl:2019 185 04 55 24 New value set for watermark SUPPLY.VOLTAGE.HIGH[-5VSEC]: -4.693947 Volts.123B18619.194ALERT#19404:21:221-1

Instrument constant changes for the week before 2019186 (3 sec, 0 rows)

SL burned out? (AP results when SL is ON) (0 sec, 17 rows)

Non-critical warnings from b-files' comments for days between 2019186 and 2019186 (0 sec, 7 rows)

TypeCommentcountpriorityfnamelocationserialtimevalueminmax
rl:2019 185 04 25 19 Reset Warm -> Tepid from 2019 185 04 25 14 until 2019 185 04 25 14 . Counts: 1,013B18619.194ALERT#19404:21:191-1
ds:DS intensity too high ( 209505/cy) quitting.399B18619.194ALERT#19408:52:191-1
ds:Suppressed oscillation of filter wheel.45B18619.194ALERT#19412:50:201-1
zs:HOME key pressed199B18619.194ALERT#19415:34:451-1
zs:ZS measurement terminated15B18619.194ALERT#19415:34:481-1
zs:Going to menu199B18619.194ALERT#19415:34:501-1
si:New SI, HC = -16, NC = -805 Old = -16 , -805199B18619.194ALERT#19415:43:561-1

Measurements and tests summary.


_jday_loc_serial_skcaodedsdtedhghprssluvzs
186ALERT#194ecsc1171191145168226
(0 sec)

Relative and absolute humidity for day 2019186 (3 sec, 1 rows)

SerialRH, %AH, g/m37-day RH historyDes-nt replcd days ago
1943.20.6