Database last updated on Fri, 11 Sep 2026 21:40:55 +0000 by localhost
Critical warnings from b-files' comments for days between 2019167 and 2019167 (0 sec, 2 rows)
| Type | Comment | count | priority | fname | location | serial | time | value | min | max |
| rl: | 2019 166 04 57 32 New value set for watermark SUPPLY.VOLTAGE.HIGH[-5VSEC]: -4.688570 Volts. | 13 | 3 | B16719.194 | ALERT | #194 | 04:15:11 | | 1 | -1 |
| ds: | Suppressed oscillation of filter wheel. | 5 | 5 | B16719.194 | ALERT | #194 | 11:36:18 | | 1 | -1 |
Instrument constant changes for the week before 2019167 (4 sec, 0 rows)
SL burned out? (AP results when SL is ON) (0 sec, 13 rows)
Non-critical warnings from b-files' comments for days between 2019167 and 2019167 (0 sec, 3 rows)
| Type | Comment | count | priority | fname | location | serial | time | value | min | max |
| rl: | 2019 166 04 20 01 Reset Warm -> Tepid from 2019 166 04 19 55 until 2019 166 04 19 55 . Counts: 1,0 | 1 | 3 | B16719.194 | ALERT | #194 | 04:15:09 | | 1 | -1 |
| rl: | 2019 167 06 02 04 New value set for watermark SUPPLY.VOLTAGE.HIGH[-5VSEC]: -4.570281 Volts. | 4 | 3 | B16719.194 | ALERT | #194 | 04:15:32 | | 1 | -1 |
| sc: | SC intensity too low ( 436/cy), skipping. | 2 | 3 | B16719.194 | ALERT | #194 | 08:30:33 | | 1 | -1 |
Measurements and tests summary.
| _jday | _loc | _serial | _skc | aode | ds | dt | ed | hg | hp | rs | sl | uv | zs |
| 167 | ALERT | #194 | ecsc | 111 | 111 | 1 | 1 | 6 | 6 | 1 | 6 | 86 | 28 |
(0 sec)
New operators? (0 sec, 29 rows)
| Operator | Serial |
| (HH:MM) | 194 |
Relative and absolute humidity for day 2019167 (2 sec, 1 rows)
| Serial | RH, % | AH, g/m3 | 7-day RH history | Des-nt replcd days ago |
| 194 | 9.0 | 1.4 |