Critical warnings from b-files' comments for days between 2023186 and 2023186 (0 sec, 1 rows)
| Type | Comment | count | priority | fname | location | serial | time | value | min | max |
| rl: | 2023 186 18 51 31 Motor #9: appears stuck with a reference sensor activated. | 2 | 1 | | | #192 | 06:06:44 | | 1 | -1 |
Instrument constant changes for the week before 2023186 (2 sec, 0 rows)
HG burned out? (AP results when HG is ON) (0 sec, 10 rows)
Non-critical warnings from b-files' comments for days between 2023186 and 2023186 (0 sec, 8 rows)
| Type | Comment | count | priority | fname | location | serial | time | value | min | max |
| rl: | 2023 186 18 51 31 Motor #9: appears stuck with a reference sensor activated. | 1 | 3 | | | #192 | 06:06:44 | | 1 | -1 |
| rl: | 2023 186 18 51 31 Motor #10: appears stuck with a reference sensor activated. | 1 | 3 | | | #192 | 06:06:47 | | 1 | -1 |
| rl: | 2023 186 18 53 34 Bad command string. Command string buffer cleared. | 1 | 3 | | | #192 | 06:06:49 | | 1 | -1 |
| rl: | 2023 186 18 53 34 Bad command string. Command string buffer cleared. | 1 | 3 | | | #192 | 06:06:49 | | 1 | -1 |
| rl: | 2023 186 18 53 34 < | 1 | 3 | | | #192 | 06:06:51 | | 1 | -1 |
| ds: | Suppressed oscillation of filter wheel. | 1 | 5 | | | #192 | 13:40:04 | | 1 | -1 |
| ds: | HOME key pressed | 3 | 99 | | | #192 | 18:07:51 | | 1 | -1 |
| pd: | Going to menu | 3 | 99 | | | #192 | 18:08:45 | | 1 | -1 |
Measurements and tests summary.
| _jday | _loc | _serial | _skc | aode | ds | dt | ed | hg | hp | rs | sl | uv | zs |
| 186 | EUREKA | #192 | 188 | 188 | 2 | 2 | 16 | 15 | 2 | 7 | 41 | 8 |
(0 sec)
Relative and absolute humidity for day 2023186 (2 sec, 1 rows)
| Serial | RH, % | AH, g/m3 | 7-day RH history | Des-nt replcd days ago |
| 192 | 2.4 | 0.4 |