Database last updated on Sat, 24 Jan 2026 20:40:55 +0000 by localhost

Critical warnings from b-files' comments for days between 2025247 and 2025247 (0 sec, 3 rows)

TypeCommentcountpriorityfnamelocationserialtimevalueminmax
rl:2025 246 05 35 13 New value set for watermark SUPPLY.VOLTAGE.HIGH413B24725.191TORON#19105:31:321-1
rl:2025 247 17 04 33 PMT counter failure11B24725.191TORON#19105:33:211-1
ds:Suppressed oscillation of filter wheel.75B24725.191TORON#19116:38:271-1

Instrument constant changes for the week before 2025247 (2 sec, 0 rows)

Non-critical warnings from b-files' comments for days between 2025247 and 2025247 (0 sec, 2 rows)

TypeCommentcountpriorityfnamelocationserialtimevalueminmax
rl:2025 246 05 34 12 Reset Warm -> Tepid from 2025 246 05 34 06 until 2025 246 05 34 06 . Counts: 1,013B24725.191TORON#19105:31:291-1
re:Zenith discrepancy = -11199B24725.191TORON#19105:35:29-11-1010

Measurements and tests summary.


_jday_loc_serial_skcaodedsdtedfmfzhghprsslzs
247TORONTO#191ecnouv,Next schedule will be ecnouv (now ecnouv ZS= 93 ZS= 93.064)64712219822192759
(0 sec)

Relative and absolute humidity for day 2025247 (2 sec, 1 rows)

SerialRH, %AH, g/m37-day RH historyDes-nt replcd days ago
19121.95.0