Database last updated on Tue, 27 Jan 2026 20:10:53 +0000 by localhost
Critical warnings from b-files' comments for days between 2025050 and 2025050 (0 sec, 1 rows)
| Type | Comment | count | priority | fname | location | serial | time | value | min | max |
| rl: | 2025 050 05 44 54 New value set for watermark SUPPLY.VOLTAGE.HIGH | 29 | 3 | B05025.191 | TORON | #191 | 05:37:25 | | 1 | -1 |
Instrument constant changes for the week before 2025050 (2 sec, 0 rows)
SL burned out? (AP results when SL is ON) (0 sec, 2 rows)
HG burned out? (AP results when HG is ON) (0 sec, 1 rows)
Non-critical warnings from b-files' comments for days between 2025050 and 2025050 (0 sec, 4 rows)
| Type | Comment | count | priority | fname | location | serial | time | value | min | max |
| rl: | 2025 050 05 43 38 Reset Warm -> Tepid from 2025 050 05 43 33 until 2025 050 05 43 33 . Counts: 1,0 | 1 | 3 | B05025.191 | TORON | #191 | 05:37:23 | | 1 | -1 |
| re: | Zenith discrepancy = -11 | 1 | 99 | B05025.191 | TORON | #191 | 05:45:09 | -11 | -10 | 10 |
| fz: | Suppressed oscillation of filter wheel. | 3 | 5 | B05025.191 | TORON | #191 | 13:40:12 | | 1 | -1 |
| ds: | DS intensity too high ( 211264/cy) quitting. | 2 | 99 | B05025.191 | TORON | #191 | 14:11:15 | | 1 | -1 |
Measurements and tests summary.
| _jday | _loc | _serial | _skc | aode | ds | dt | ed | fm | fz | hg | hp | rs | sl | zs |
| 050 | TORONTO | #191 | ecnouv,Next schedule will be ecnouv (now ecnouv ZS= 93 ZS= 93.86901) | 71 | 73 | 2 | 2 | 21 | 8 | 14 | 13 | 2 | 8 | 50 |
(0 sec)
Relative and absolute humidity for day 2025050 (2 sec, 1 rows)
| Serial | RH, % | AH, g/m3 | 7-day RH history | Des-nt replcd days ago |
| 191 | 11.9 | 1.8 |